Operators Manual
D3740 25
Setting/
Display
D.A.N Options
(
Bold
= Default)
Description
MENU 3
Display
Count
(
COUNT
COUNT
COUNT
COUNT
)
301
.01,
.02
,.05,.1,.2,.5,
1,2,5,10,20, 50,100
Count set too small, readings
unstable and indicator not
accurate
Auto-Range
(
ARANGE
ARANGE
ARANGE
ARANGE
)
302
ON/
OFF
Scale increases display count
size for weights over 300
again at 600 lbs/kgs.
Display Unit
(
LB
LB
LB
LB----KG
KG
KG
KG
)
303
LB
/KG
When changing weight unit
calibration adjusted so scale
displays accurately in new
display unit
Capacity
(
CAP
CAP
CAP
CAP
)
304
4000
Enter maximum weight
measurable
WM1
ADJUST
(WMA 1-1)
305
10
Increase number to smooth
weighing
WM1
ADJUST 2
(
WMA 1
WMA 1
WMA 1
WMA 1----2222
)
306
0,1,2,3,
4
,5,6,7,8,9
0=OFF Use values less than
WMA 1-1 for quick weight
response.
WM1
ADJUST 3
(
WMA 1
WMA 1
WMA 1
WMA 1----3333
)
307
400
Enter weight to activate quick
weight response.
WM2
ADJUST 1
(
WMA
WMA
WMA
WMA 2222----1111
)
311
30
Increase number to smooth
weighing
WM2
ADJUST 2
(
WMA
WMA
WMA
WMA 2222----2222
)
312
10
0=OFF Use values less than
WMA2-1 for quick weight
response.
WM2
ADJUST 3
(
WMA
WMA
WMA
WMA 2222----3333
)
313
400
Enter weight activate quick
weight response.
MENU 4
Buzzer
(
BUZZER
BUZZER
BUZZER
BUZZER
)
404
1
-4, ON, OFF
Decrease # shortens alarm
horn
Relay
(RELAY)
406
SETPNT
/OFF
Set control for 12 VDC alarm
Press enter @ setpnt gives
option to enter value
Содержание SW4600EID
Страница 46: ...SW4600EID 46 D3740...