Caution

Digi-Pas  device is calibrated in the factory for maximum 

accuracy. User can re-calibrate when necessary.

Please refer to www.digipas.com for video demonstration 

on how to calibrate the device.

®

Calibration Instruction

Note: 

Ensure the device is kept still during calibration 

Clip the device to a spirit level, make sure the device is 

securely cliped on the level, turn on the device then press 

‘CAL’ button once  the screen will display flashing “CAL1”.

Place the device on a flat and clean surface, press 

“MODE” button once, the screen will display countdown, 

wait until screen countdown reach “0” and show “CAL2”.

Rotate the device (180°) parallel to the surface then press 

“MODE” button again, the screen will display countdown, 

wait until screen countdown reach “0” and show “CAL3”.

Position the device vertically (-90°) on a level surface, 

press “MODE” button again, the screen will display 

countdown, wait until screen countdown reach “0” and 

show “CAL4”.

Rotate the device 180° parallel vertically to the surface 

(+90°) then press “MODE” button again, the screen will 

display countdown, wait until screen countdown reach “0”.

Calibration Done.