14
7.4.6 Command: L<CR> (4ch 0dh)
Response: <LF> H1H2H3 <CR><ETX>
If Hold function can be activated, it will enable/disable hold (lock) function, like the UNIT/HOLD key is
pressed, and returns scale status.
7.4.7 Command: X<CR> (58h 0dh)
Response: power off the scale, just like press down the ON/OFF/ZERO key to turn off the scale.
7.4.8 Command: all others
Response: <LF>? <CR><ETX>
Unrecognized command
7.5 Output status bit meaning:
The status bit definition:
Bit
Byte 1 (H1)
Byte 2 (H2)
Byte 3 (H3)
0
0=stable
0= not under capacity 01=normal work mode
10= hold work mode
00=not define
11= not define
1= not stable
1= under capacity
1
0= not at zero point
0= not over capacity
1= at zero point
1= over capacity
2
0=not AD over
0=not Zero Over
0= gross weight
1=AD over
1=Zero Over
1= net weight
3
0= eeprom OK
0=not Zero down
0=not AD down
1= eeprom error
1= Zero down
1=AD down
4
always 1
always 1
always 1
5
always 1
always 1
always 1
6
always 0
always 1
always 0
7
parity
Parity
parity
8.
Symbol Definitions:
0
ˉ ˉ ˉ ˉ Zero point is over the setting range
0
_ _ _ _
Zero is below the setting range
Ad
ˉ ˉ ˉ
ADC is over max. range;
Ad
_ _ _
ADC is below min. range;
ˉ ˉ ˉ ˉ ˉ
Weight signal is too large
_ _ _ _ Weight signal is too small
EEP.E0
The EEPROM can’t be accessed;
EEP.E1
The parameters are not same with backup data;
EEP.E2
The setting parameter(s) is not in normal range;.
CAL-P
x
Scale’s calibration point;
CAL.Er
There is an error in calibration
Hold(H) Hold function is active.
Net
◄
The display reading is net weight
Zero
◄
The scale is at zero point
CAP.
--
The the setting full capacity will be displayed