Dickson TH8P0 Скачать руководство пользователя страница 6

Dickson TH8P Manual V14 052418 

7.

 

Each time P

P

P

Pen 

en 

en 

en H

H

H

Home

ome

ome

ome is pressed the two pens will switch positions (moving the most 

recently adjusted position) until On/Off

On/Off

On/Off

On/Off is pressed to exit pen adjust mode.   

 
Note: Pressing the On/Off button will exit the adjustment and return the unit to normal 
operation. Each time the Pen Home or On/Off button is pressed, the adjustment currently in 
progress will be stored. The unit will take one minute to exit Pen Adjust Mode once the On/Off 
button is pressed. 

 

 
C

C

C

CALIBRATION

ALIBRATION

ALIBRATION

ALIBRATION    

The R200/R250 Replaceable Sensor was carefully tested and calibrated before being shipped 
from the factory. For greatest accuracy, we recommend replacing the sensor every 6-12 months 
(does not include N400 and N450). 
 
To order a Calibrated Replacement Sensor Call customer service at (630) 543-3747 or go to 
www.dicksondata.com. 
 
When your new Replaceable Sensor arrives, simply turn off the recorder, remove and discard the 
old sensor, plug in the new one and power the recorder back on.  Your recorder will continue to 
record temperature and humidity without interruption. 
    
USER CALIBR

USER CALIBR

USER CALIBR

USER CALIBRATION

ATION

ATION

ATION    

If you have an accurate standard to compare against, the Replaceable Sensor calibration can be 
adjusted at one point.  This will not adjust the span and is not as accurate as replacing the 
Replaceable Sensor with a newly calibrated one. 
 
1.

 

To activate Calibration mode, turn the unit on and press and hold both the On/Off

On/Off

On/Off

On/Off and the 

Adjust

Adjust

Adjust

Adjust----Down

Down

Down

Down button until the LED is solid green. The LED will then blink amber at which 

point only the pen being adjusted will show on the display.  
Note:

Note:

Note:

Note:

 

Make sure to press the Adjust-Down button first so the unit will not turn off 

 

2.

 

To raise the unit of measurement, press the Adjust

Adjust

Adjust

Adjust----Down

Down

Down

Down button. To lower the unit of 

measurement, press the Adjust

Adjust

Adjust

Adjust----Up

Up

Up

Up button. Pressing Pen home

Pen home

Pen home

Pen home switches between the red and 

blue pens on a two pen unit and stores the current adjustment value.  

 

3.

 

When Calibration is complete, press the On/Off

On/Off

On/Off

On/Off button. The adjustment is stored in memory 

even after you turn the unit off or if AC power fails. 

 

Note:

Note:

Note:

Note: After two hours, if no buttons are pressed, the unit will time out of Calibration mode 
and resume normal operation. If you wish to cancel the Calibration, simply enter Calibration 
mode and toggle through the steps without adjusting displayed readings. Exit by pressing 
the On/Off

On/Off

On/Off

On/Off button. You have now restored factory calibration settings. 

 

Note: 

Note: 

Note: 

Note: It is recommended that you use a controlled chamber when determining if Calibration 
is necessary.  Adjusting the pen locations in an open room is not recommended as humidity 
can vary greatly within a very small area. 

 
    
NEED T

NEED T

NEED T

NEED TO KNOW

O KNOW

O KNOW

O KNOW    

The pen movement for the TH8P0 is slower than on the AC powered models in order to conserve 
battery power.  In 24-hour mode the pen will move every 1 minute and 30 seconds.  In 7-day 
mode the pen will move every 8 minutes and 30 seconds. 
 

Содержание TH8P0

Страница 1: ...a Set the chart time manually by inserting a coin into the groove in the chart hub and turning clockwise until the correct hour and day if applicable on the chart is referenced to the timing arrow jus...

Страница 2: ...ed on the outside edge of the chart press the Pen Home Pen Home Pen Home Pen Home to move pen s to recording position If the pens are located on the chart press the Pen Home Pen Home Pen Home Pen Home...

Страница 3: ...n will scroll through the five options until the A A A Alarm larm larm larm button is pressed to exit from alarm adjust Each press of the Pen Home Pen Home Pen Home Pen Home or Alarm Alarm Alarm Alarm...

Страница 4: ...Use a Pen or small screw driver to flip the switches Remember to install the correct chart to match the corresponding Dip Switch setting Re Re Re Recording Time cording Time cording Time cording Time...

Страница 5: ...s over time 1 While the unit is on press and hold both Pen Home Pen Home Pen Home Pen Home and On Off On Off On Off On Off buttons until the LED is solid green Release the Pen Pen Pen Pen Home Home Ho...

Страница 6: ...th the On Off On Off On Off On Off and the Adjust Adjust Adjust Adjust Down Down Down Down button until the LED is solid green The LED will then blink amber at which point only the pen being adjusted...

Страница 7: ...al Why does the display read PROB Why does the display read PROB Why does the display read PROB Why does the display read PROB Check the TH sensor pod Make sure it is properly plugged into the unit Is...

Страница 8: ...won t the unit power up Remove the batteries and power adapter for a minute or two this will reset the unit The unit should respond when the adapter is plugged in WARRANTY WARRANTY WARRANTY WARRANTY D...

Страница 9: ...The Replaceable Sensor must be connected to the Instrument for the Instrument to operate properly Calibration Calibration Calibration Calibration Only the Replaceable Sensor is calibrated The sensor...

Отзывы: