 Why don't the display and chart match?

Why don't the display and chart match?

Why don't the display and chart match?

Why don't the display and chart match?  

 

Dip switches set for a specific range, but using chart for another range, or vice versa.  

 

Pens(s) not inserted on pens arm(s) all the way  

 

To adjust pens(s) to match chart see “Pens Home Adjust” in the “Need to Know” section 
of this manual. 

 Why does the display read PROB?

Why does the display read PROB?

Why does the display read PROB?

Why does the display read PROB?  

 

Check the sensor pod.  Make sure it is properly plugged into the unit. 

 

Is the unit locked up? This can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the unit 
may appear to be working, but the readings won't change, also the chart will not rotate. 
Remove power and battery, and then re-power. 

 Why does calibration seem to be off?

Why does calibration seem to be off?

Why does calibration seem to be off?

Why does calibration seem to be off?  

 

What is the tolerance of the unit it's being compared to?  

 

It's ok if the unit is within the sum of the two tolerances.  

 

Has an outside calibration house attempted calibration? It may not have been adjusted 
properly.  

 

Walk through calibration adjustment, found in the manual 

 Why won't the battery back

Why won't the battery back

Why won't the battery back

Why won't the battery back----up work?

up work?

up work?

up work?  

 

Are there good batteries in the Chart Recorder? 

 

Keep in mind that battery backup varies significantly depending on temperature, pens 
movement and chart rotation speed.  

 

If there was a momentary power outage (Brown out), the unit may not have had enough 
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or 
shut off in this situation) This situation may also occur if the unit is plugged into an outlet 
that shares the circuit with other machinery that has phase motors or compressors that 
cycle periodically. As these other machines cycle, they momentarily have a high current 
draw, therefore drawing from the unit. 

  Why won't the chart recorder respond to changes?

Why won't the chart recorder respond to changes?

Why won't the chart recorder respond to changes?

Why won't the chart recorder respond to changes?  

 

Is the unit locked up?  This can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the unit 
may appear to be working, but the readings won't change, also the chart will not rotate. 
Remove power and batteries, and then re-power.  
 

  Why won't the unit power up?

Why won't the unit power up?

Why won't the unit power up?

Why won't the unit power up?  

 

Remove the batteries and power adapter for a minute or two, this will reset the unit. The 
unit should respond when the adapter is plugged in. 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

www.

.com

[email protected]

1.800.561.8187

Содержание TH6P1

Страница 1: ...rt 4AA batteries and plug in AC adapter See Figure 3 Unit will power on 4 4 4 4 The instrument will move pens to the appropriate readings 5 5 5 5 Install the chart that matches selected Dip Switch set...

Страница 2: ...tions Keypad and Button Functions Keypad and Button Functions On Off On Off On Off On Off The On Off key turns the unit on and off Pen Pen Pen Pen Home Home Home Home If the pens are located on the ou...

Страница 3: ...ve options until the A A A Alarm larm larm larm button is pressed to exit from alarm adjust Each press of the Pen Pen Pen Pen Home Home Home Home or Alarm Alarm Alarm Alarm button will store the new s...

Страница 4: ...Home Pen Home Pen Home first so the unit will not turn off 2 Both pens will move to the outer edge of the chart and then the shorter red red red red pen will move to the outer chart ring maximum temp...

Страница 5: ...complete press the On Off On Off On Off On Off button The adjustment is stored in memory even after you turn the unit off or if AC power fails Note Note Note Note After two hours if no buttons are pr...

Страница 6: ...ck Why won t the battery back Why won t the battery back Why won t the battery back up work up work up work up work Are there good batteries in the Chart Recorder Keep in mind that battery backup vari...

Страница 7: ...nsors will work with the Instrument models listed No other Instrument will work with the Replaceable Sensors listed The Replaceable Sensor must be connected to the Instrument for the Instrument to ope...

Отзывы: