Dickson KT6P Manual V3 

Pen 1 alarm minimum 
Pen 1 alarm maximum 

 

3.

 

In order to set pen alarm minimum and maximum, pressing Adjust-Up will increase the alarm value, 
and pressing Adjust-Down will decrease the alarm value. There is acceleration if the Adjust-Up button 

is held down. 

 

Repetitively pressing the Pen Home button will scroll through the three options until the Alarm 
button is pressed to exit from alarm adjust. Each press of the Pen Home or Alarm button will store the 

new settings. The unit will take one minute to exit Alarm Set Mode once the Alarm button is pressed. 
 

4.

 

If the alarm is triggered, the LED will show as solid red and the alarm will sound. Press the Alarm 

button to silence the audible alarm.  

 
Alarm Relays (KT6P5) 

The SPST 24V 500mA relay contacts are normally open and will close on alarm conditions when the 
alarms are enabled. Relay contacts are always functional when the alarm is enabled. The relay will close 
only during minimum and maximum alarm conditions. Relays will not operate under battery only power 
conditions

DIP SWITCH SETUP 
To setup the KT6 recorder for your specific application, you might need to change some of the Dip 

Switches. The Dip Switches are located on the back of the unit.  (Figure 3) 
Use a pen or small screw driver to flip the switches.  Remember to install the correct chart to match 
the corresponding Dip Switch setting. 

 

Recording Time: The KT6 recorder has two different recording time options 
7 Day 

#2 Down 

24 Hour 

#2 Up 

F/C: Record in degrees F or C 

#3 Down 

#3 Up 

Temperature Range: Select the temperature ranges that matches the chart paper  
0 to +100 F/C 

#4 Down 
#5 Down 
#6 Down 

-50 to 0 F/C 

#4 Down 

#5 Down 
#6 Up 

-100 to 0 F 

#4 Down 
#5 Up 
#6 Up 

-50 to 50 F/C 

#4 Up 

#5 Down 
#6 Down 

0 to 50 F/C 

#4 Down 
#5 Up 

www.

.com

[email protected]

1.800.561.8187

Содержание KT6P1

Страница 1: ...he timing arrow just to the right of the pen tip on the dial Press Pen Home to move the pen s back onto the chart The following feature should be used for fine adjustments only b To adjust the chart t...

Страница 2: ...set the alarm make sure unit is on and press and hold the Alarm button The LED will blink red rapidly for about five seconds then the LED will turn solid green Release the Alarm button and the LED wi...

Страница 3: ...ill close on alarm conditions when the alarms are enabled Relay contacts are always functional when the alarm is enabled The relay will close only during minimum and maximum alarm conditions Relays wi...

Страница 4: ...d green Release the Pen Home and On Off buttons The LED will flash amber and green for one second then the LED will turn off Note Press Pen Home first so the unit will not turn off 2 The pen will move...

Страница 5: ...tion If you wish to cancel the Calibration simply enter Calibration mode and toggle through the steps without adjusting displayed readings Exit by pressing the On Off button You have now restored fact...

Страница 6: ...d enough time to recognize this and switch to battery mode The Chart Recorder may lock up or shut off in this situation This situation may also occur if the unit is plugged into an outlet that shares...

Страница 7: ...t to operate properly Calibration Only the Replaceable Sensor is calibrated The sensor and all calibration defaults and adjustments are stored on the Replaceable Sensor Accuracy is strictly controlled...

Отзывы: