Figure 1 

Grain Hopper Sensors

 

 

INTERNAL CLEANING 

Performing continuous tests can result in material accumulation around 

the critical internal components and adversely affect the measurement. 
Two types of cleaning are recommended on an as needed basis: 

• Daily clean 

• Extensive clean (see Operator’s Manual for instructions) 

DAILY CLEAN METHOD 

A daily clean method allows cleaning the cell and door using an 

automated process. During the cleaning sequence, the hopper door 

automatically opens.  

To start the cleaning process: 

1.  Select 

Device Information

 from the Main Menu. 

2.  Press 

  A pop up message will appear with the words  

“Cleaning Mode is Active”.  

3.  Remove the grain drawer. 
4.  Using the supplied brush, manually remove any loose or stuck grain 

or dust from the measuring cell. 

5.  Press the 

CLOSE

 button to return instrument to normal operation. 

 

Hands should be clear from 

inside the instrument before 

pressing the CLOSE button. 

6.  Insert grain drawer. 

Press 

 to return to the Home Screen. 

Содержание GAC 2700-UGMA

Страница 1: ...Quick Start Guide GAC 2700 UGMA Grain Analysis Computer...

Страница 2: ...47 CFR Part 18 109 c ISM Class A and ANSI C63 4 2014 and Canada ICES 001 CISPR11 Group 2 Class A The GAC 2700 product series is exempted from the Radio Equipment Directive 2014 53 EU as the RF signal...

Страница 3: ...rument and its sensors on a regular and as needed basis depending on the amount of dust dirt and debris encountered in the instrument s use by monitoring performance using daily check samples and by a...

Страница 4: ...he bag the packaging material and the carton to use for future transport of the instrument 3 Ensure the installation location is level using the bubble level at the top of the instrument adjust the fe...

Страница 5: ...bracket on the left side wall 6 Press down on the bracket to release 7 Slide the bracket toward the bottom of the instrument 8 Insert the hex bolt into the operating location hold and tighten using th...

Страница 6: ...instrument between imperial and metric units The following steps will be displayed during the Out of Box Experience 1 Highlight the desired Language and press Save 2 Press the OK button 3 Highlight t...

Страница 7: ...ailed explanation on the above configuration settings C o n d u c t i n g G r a i n An a l y s e s The GAC 2700 is designed to provide minimal user interaction to run grain moisture measurements From...

Страница 8: ...horized service center For customers that require a more extensive cleaning contact and schedule your instrument for cleaning with your dealer or authorized service center Note The following recommend...

Страница 9: ...aning the cell and door using an automated process During the cleaning sequence the hopper door automatically opens To start the cleaning process 1 Select Device Information from the Main Menu 2 Press...

Страница 10: ...o Low 13 No Communication 14 Instrument High Temp Limit Exceeded 15 Unit to Grain Differential 16 Internal Power Supply Out of Spec 17 Unable to Predict Moisture 18 Pre Analysis Timeout 19 Instrument...

Страница 11: ...from misuse neglect accident or improper installation or maintenance any expenses or liability for repairs made by outside parties without DICKEY john s written consent damage to any associated equipm...

Страница 12: ...the United States and may be protected under other country s trademark registrations 6015415 Rev C A D I V I S I O N O F T S I 5200 Dickey John Road Auburn IL 62615 www dickey john com 1 217 438 3371...

Отзывы: