OPERATOR’S MANUAL

Dj F-2000 Hay Moisture Meter

11001-1355-200511

OPERATION 5

OPERATION

CHANGING SET-POINT

To change the set-point value, press the 

Set-Point Increase

 button. The 

meter will display the current set-point. To scroll forward to a higher value, 
hold the 

Set-Point Increase

 button while the current value is displayed and 

scroll to the set-point value desired.

To scroll backward through the set-point values, press and release the 

Set-Point Increase

 button. Within one second, press and hold the 

Set-Point Decrease

 button. Continue to hold the 

Set-Point Decrease 

button and the set-point will decrease. 

When scrolling in either direction, release the button to stop at the desired 
set-point. If the meter reads a %MC (percent moisture content value) higher 
than that of the set-point, a buzzer will sound.

CHECKING ACCUMULATED READINGS

This feature allows the user to view the total number of all accumulated 
readings, the average of those readings, and the highest stored reading.

To add a reading to the sum of all previously stored readings, press and 
release the 

Read

 button within 2 seconds. Pressing and holding the 

Read

 

button will repeat the read cycle but will not add a new reading to the 
storage until the button is released.

To view the readings, press and release the 

Check 

button. The meter will 

display the number of accumulated readings for one second, then the 
average of those readings for two seconds. It will then display the highest 
stored reading for two seconds. The total cycle time is five seconds.

To erase all accumulated readings, hold the 

Check 

button for more than 

five seconds until the meter displays “0”.

To keep the accumulated readings in memory, release the 

Check

 button 

before the total cycle time is complete. The meter will accumulate up to 100 
readings. After all 100 readings are stored, the meter will not add new 
readings until the memory has been cleared. It will also continue to display 
the average of all 100 readings as a reminder that the memory is full.

Readings below 6% will be displayed as “.0”. Those above 40% will be 
displayed as “99.9”. Neither will be added to the accumulated readings or 
used in the calculation of average or highest reading.

CHECKING CALIBRATION

Remove the probe from the top of the meter. Press and hold the 

Read

 

button and 

Check

 button simultaneously. The meter is in calibration if it 

displays “12.0” (+/- 0.2) on the scale.

Set-point Increase button

Set-point Decrease button

Read button

Check button

Содержание Dj F-2000

Страница 1: ...3 Operation 5 Changing Set Point 5 Checking Accumulated Readings 5 Checking Calibration 5 Resetting The Meter 6 Testing Baled Hay 6 Testing In The Windrow 6 Factors Affecting Readings 9 Curing 9 Dens...

Страница 2: ...Dj F 2000 Hay Moisture Meter 11001 1355 200511 ii TABLE OF CONTENTS...

Страница 3: ...llows Be alert to the possibility of personal injury or death Use of the word WARNING indicates a potentially hazardous situation which if not avoided could result in death or serious injury Use of th...

Страница 4: ...OPERATOR S MANUAL Dj F 2000 Hay Moisture Meter 11001 1355 200511 2 SAFETY NOTICES...

Страница 5: ...ese factors are critical to the hay producer s main goal to produce better quality hay and minimize economic losses FEATURES The Dj F 2000 offers the following features 6 40 moisture range for alfalfa...

Страница 6: ...OPERATOR S MANUAL Dj F 2000 Hay Moisture Meter 11001 1355 200511 4 INTRODUCTION Figure 1 Dj F 2000 Hay Moisture Meter Set point Increase button Set point Decrease button Read button Check button...

Страница 7: ...button will repeat the read cycle but will not add a new reading to the storage until the button is released To view the readings press and release the Check button The meter will display the number o...

Страница 8: ...veral different parts of the bale and the highest readings used as a guideline The arrangement and compaction of hay fibers in a bale may have an effect on meter readings When testing high density bal...

Страница 9: ...prod to the hay in the windrow and press down on the electrode handle until the head of the screw touches the palm of your hand Press the Read button and take a reading Perform several tests on the h...

Страница 10: ...OPERATOR S MANUAL Dj F 2000 Hay Moisture Meter 11001 1355 200511 8 OPERATION...

Страница 11: ...ill be The more curing that has been allowed to take place the greater uniformity in moisture distribution can be expected The validity of the meter readings is closely related to the care spent in sa...

Страница 12: ...alitative indication of high moisture content DICKEY john moisture meters use the relationship existing between electrical conductivity and moisture content in hay As moisture content increases so doe...

Страница 13: ...to note the high readings and the frequency at which they occur USE OF PRESERVATIVES Hay preservatives or stabilizers may also have an affect on meter readings Normally a bale of hay treated with pres...

Страница 14: ...OPERATOR S MANUAL Dj F 2000 Hay Moisture Meter 11001 1355 200511 12 FACTORS AFFECTING READINGS...

Страница 15: ...r in a clean dry place The optional protective carrying case in an ideal storage place when the meter is not in use Change the 9 volt battery as needed Continued use with a low battery may cause the m...

Страница 16: ...OPERATOR S MANUAL Dj F 2000 Hay Moisture Meter 11001 1355 200511 14 METER CARE...

Страница 17: ...m date of original installation and is returned to DICKEY john within 30 days after such defect is discovered DICKEY john will at our option either replace or repair said part This warranty does not a...

Отзывы: