Omini o
perator’s
manual
version :19.04.17
18
Change of RESIN CARTRIDGE
1.
Close the tap water hand valve.
2.
Close the storage tank hand valve.
3.
Open the water purifier exit hand valve to depressurize the network; the
indicator SOURCE » of the electronic unit lights. Leave the water purifier exit
hand valve. Turn off the water purifier (switch in position « 0 »), then unplug the
power plug.
4.
Remove the two glass doors.
5.
Using the key (supplied with water purifier) carefully unscrew (turn to the left)
the transparent filter holder
6.
Remove the filter cartridge. Rinse the filter holder. Introduce the new cartridge
resin cartridge
included in kit ref. 950020
7.
WARNING: the filter is filled with water, a mop at hand is recommended!
8.
Check that the O-ring is properly positioned in the filter holder groove, then re-
screw it. A « good tightening by hand » is enough.
9.
Dry the floor at the bottom of the water purifier in order to detect a possible
leak during the starting up.
10.
Open the tap water hand valve.
11.
Re-plug the power plug then start up the water purifier. The indicators
« POWER », « PUMP » and « FLUSH » light: the water purifier starts 2 minutes of
rinsing cycle of the osmosis membrane. The rinsing is ended when the indicator
« FLUSH » switch off; then the water purifier is on production.
12.
Open the storage tank hand valve.Let filling all filters holder. After some time,
the water must flow at the water purifier exit: let flowing 10 to 20 liters of water
so as the drain
13.
Close the water purifier exit hand valve. Put back glasses doors. Then the water
purifier is operational.