Change of the POST TREATMENT CARTRIDGE: SEDIMENT 1 µm  

6.4.6

 

1.

 

Close the tap water hand valve.

 

2.

 

Close the storage tank hand valve.

 

3.

 

Open  the  water  purifier  exit  hand  valve  to  depressurize  the  network;  the  indicator 
« SOURCE » of the electronic unit lights. Leave the water purifier exit hand valve. Turn off the 
water purifier (switch in position « 0 »), then unplug the power plug.

 

4.

 

Remove the two glass doors.

 

5.

 

Using  the  small  key  (supplied  with  water  purifier)  carefully  unscrew  (turn  to  the  left)  the 
transparent filter 

 

6.

 

WARNING: the filter is filled with water, a mop at hand is recommended! 

7.

 

Remove the filter cartridge. Rinse the filter holder. Place the new cartridge « Sediment Filter 
1 µm 10” 

included in kit ref. 950019

 

8.

 

Check that  the  O-ring  is  properly  positioned  in  the  filter  holder  groove,  then re-screw  it.  A 
« good tightening by hand » is enough. 

9.

 

Dry the floor at the bottom of the water purifier in order to detect a possible leak during the 
starting up. 

10.

 

Open the tap water hand valve. 

11.

 

Re-plug the power plug then start up the water purifier. The indicators « POWER », « PUMP » 
and  « FLUSH »  light:  the  water  purifier  starts  2  minutes  of  rinsing  cycle  of  the  osmosis 
membrane.  The  rinsing  is  ended  when  the  indicator  « FLUSH »  switch  off;  then  the  water 
purifier is on production. 

12.

 

Open  the  storage  tank  hand  valve.  Let  filling  all  filters  holder.  After  some  time,  the  water 
must flow at the water purifier exit: let flowing 10 to 20 liters of water so as the drain 

13.

 

Close  the  water  purifier  exit  hand  valve. Put  back  glasses  doors. Then  the  water  purifier  is 
operational. 

 

 

Содержание O-Maxi OM Series

Страница 1: ...O Maxi operator s manual Version 2020 02 10 0 Operator Manual For Water purifier ...

Страница 2: ... Conditions 10 4 2 Connections 10 4 3 Connections of Resin bottle and Sampling system to test membrane exit 11 4 4 Connection of the Pressurized tank 11 4 5 Installation of the water purifier membrane 12 System start up 13 5 Maintenance and service 14 6 6 1 Maintenance calendar 14 Calandar for the version 1 14 6 1 1 Calendar for the version 2 14 6 1 2 6 2 Troubleshooting 15 Version 1 15 6 2 1 Vers...

Страница 3: ... the manual please get familiarized with the following icons used in this manual Electric Shock Warnings 2 2 SAFETY INSTRUCTIONS To ensure the product SECURITY and RELIABILITY all repairing must be realized with spare parts available with our after sales service If the power cable is damaged it must be replaced Unplug the water purifier power plug Don t pull on the wire Before all maintenance on t...

Страница 4: ...most effective and elaborated membranous separation process The demineralization principle by IONS EXCHANGERS RESINS These two associated technologies allow getting water with excellent quality regarding physical composition chemical composition mineral and organic and micro organic population The O Maxi water purifier exists in 2 versions VERSION 1 serial number OM xxx VERSION 2 serial number OM ...

Страница 5: ...rator s manual Version 2020 02 10 4 3 2 OVERVIEW OF THE SYSTEM FRONT VIEW Storage tank Resin Bottle Post treatment Reverse osmosis membrane Booster Pump Pressure Gauge Electronic control case Version 1 Version 2 ...

Страница 6: ...O Maxi operator s manual Version 2020 02 10 5 3 3 OVERVIEW OF THE SYSTEM BACK VIEW Filter for Sediments 5µm Block active charcoal fitler Filter for Sediments 1µm ...

Страница 7: ... used to check storage tank pressure The max value possible is about 3 to 3 5 bars 43 5 PSI when the water purifier is standby LED Power and Full switch on 3 5 GENERAL DESCRIPTION The O Maxi water purifier is a water treatment system that includes a series of prefiltration cartridges for the network water Sediment filter 5 μm Block carbon filter Sediment filter 1 µm Once this water is treated it i...

Страница 8: ...uency of around 6 hours following a pressure on the MANUAL button SOURCE indicator turned on Lack of flow or pressure on water supply FULL indicator turned on Full storage tank POWER indicator turned on Water purifier under tension this indicator is switched on permanently whatever the other indicators states are MANUAL push button Press this button makes starting a forced rinsing of the membrane ...

Страница 9: ... light ON Full storage tank POWER light ON Water purifier switched ON this light is always ON whatever the state of the other lights WASHING button Pushing this button launches a membrane rinsing during around 1 min 30 CONDUCTIVITY ALARM light Informs you about the water conductivity Water purifier producing Water purifier rinsing Lack of water supply Full storage tank Produced water conductivity ...

Страница 10: ...uction flow at 10 C 40 liters hour Resins type Mixbed ions exchange resins Resins volume 11 liters Maximum supply water temperature 38 C 100 F Maximum hardness without protection 40 f French degree of TH Admissible pH 3 to 11 Mini maxi supply pressure 2 5 bars 29 87 PSI Dimensions l h w 55 X 73 X 45 cm Indicative weight 49 kg This system is recommended for daily consumption over 50 liters ...

Страница 11: ...230 V 50 Hz EARTH Connect the device in a dry area more than two meters away Water drain with a siphon or the drain clamp supplied with the water purifier 4 2 CONNECTIONS Connect a hand valve on the purified water outlet Connect to the resin bottle outlet Connect to the resin bottle inlet Connect to the storage tank Connect to the drain Must be always OPEN to not damage the system Connect a hand v...

Страница 12: ...N OF THE PRESSURIZED TANK Inlet Resin bottle to be connected to the water purifier Outlet Resin bottle to be connected to the water purifier Outlet to test membrane conductivity IN OUT Outlet Pressurized Tank to be connected to the water purifier This valve is open in normal operating mode This valve is closed in normal operating mode ...

Страница 13: ...al keys provided with the accessories 4 Once the membrane cap is removed insert the new membrane Figure 3 peripheral seal at the top to the complete stop the end of the collecting tube must be flush with that of the membrane holder Figure 4 figure 1 figure 2 figure 3 figure 4 5 Check that the O ring is well positioned at the bottom of the retaining wall of the membrane holder see the plan 6 Unscre...

Страница 14: ...t 20 liters flow 9 Then carry out a conductivity test see paragraph 6 4 2 2 The value obtained must be 0 to 0 1 μS cm Then close the water purifier exit hand valve 10 Open the storage tank valve and wait for its full refill Full indicator lit in green 11 The control unit goes through the following phase 12 The water purifier is then operational The power indicator indicates that the osmosis unit i...

Страница 15: ...vity values Replace the resin bottle See paragraph 6 4 9 AT LEAST EVERY 6 MONTHES depending of conductivity values Replace the pretreatment filters and post treatment filter See paragraph 6 4 3 6 4 4 6 4 5 6 4 6 OCCASIONALLY Replace the reverse osmosis membrane and the flow restrictor See paragraph 6 4 7 6 4 8 Calendar for the version 2 6 1 2 Frequency Operation DAILY Check before using purified w...

Страница 16: ...tment filters before the RO membrane Weak flow of inlet water Increase the flow of the inlet water Reverse osmosis membrane clogged Change the membrane Power light ON No or low purified water production indicator PUMP light ON Pump always OFF Pump fuse burned Replace the pump fuse Light indicators FULL and PUMP alternate Defective high pressure sensor or check valve Change pressure sensor and chec...

Страница 17: ...he water inlet The inlet water pipe is bent or plugged Check the water inlet pipe POWER light switched ON Low purified water production PUMP light switded ON Pump working The pre treatment cartridge is plugged Replace the pre and post treatment filters Low water inlet flowrate Increase the water inlet flowrate The reverse osmosis membrane is plugged Replace the membrane POWER light switched ON Lac...

Страница 18: ...ing blue Pressure sensor out of order Replace the pressure sensor CONDUCTIVITY ALARM light blue Conductimeter or temperature sensor out of order Replace the sensor CONDUCTIVITY ALARM light blue yellow alternating Inverted connections between the conductimeter and temperature sensor Open the control unit and invert the connections ...

Страница 19: ...eatment filters must be replaced when a plugging significant drop of the purified water production flow and or sediments 5 µm filter seems dirty Note the 3 pre treatment cartridges average life time is from 2 to 6 months following the tap water quality Material Suspended rate measurable by the turbidity and the water purifier running time The post treatment cartridge must be replaced at the same t...

Страница 20: ...frequency of the ions exchangers resins bottle replacement Note the average life time of a reverse osmosis membrane is from 1 to 3 years following the tap water quality the water purifier running time and the preventive maintenance respect OP101 OP202 OP302 OMAXI RESIN KIT ref 950243 6 3 3 The resin bottle must be changed when the purified water conductivity increases Note the average life time of...

Страница 21: ...his theoretical flow rate represents the production at a tank outlet pressure of zero bar it decreases according to the increase in the pressure Outlet tank 6 4 1 2 Equipment 1 A graduated test tube of 500 mL 2 A chronometer 6 4 1 3 Operating method 1 Close the storage water hand valve 2 Open the water purifier exit hand valve and let flowing 30 seconds to get a stabilized flow 3 Trigger the chron...

Страница 22: ...Fill protection cap of the conductivity meter with the water to test then plunge the conductivity meter read the measured value 6 Renew the operation 2 to 3 times the retained value will be the last measurement one 7 Switch off the conductivity meter then put back the protection cap INTERPRETATION OF THE VALUE OF THE CONDUCTIVITY OF PURIFIED WATER Conductivity 1 μS cm Class 2 water according to IS...

Страница 23: ...conductivity measurement is carried out at the outlet of the storage tank connection pipe 6 4 2 3 1 Preparation 1 Close the storage tank hand valve 2 Close the ions exchange resins ENTRY hand valve 3 Remove the plug on the outlet TEST reverse osmosis membrane exit 4 Open the TEST reverse osmosis membrane exit hand valve 6 4 2 3 2 Measured 1 Switch on the instrument 2 Let the water to be tested run...

Страница 24: ...ILTERS 5 µm included in kit ref 950019 7 Check that the O ring is properly positioned in the filter holder groove then re screw it A good tightening by hand is enough 8 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 9 Open the tap water hand valve 10 Re plug the power plug then start up the water purifier The indicators POWER PUMP and FL...

Страница 25: ...ef 950019 7 Check that the O ring is properly positionned in the filter holder groove then re screw this one A good tightening by hand is enough 8 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 9 Open the tap water hand valve 10 Re plug the power plug then start up the water purifier The indicators POWER PUMP and FLUSH light the water pu...

Страница 26: ...er is filled with water a mop at hand is recommended 8 Check that the O ring is properly positioned in the filter holder groove then re screw it A good tightening by hand is enough 9 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 10 Open the tap water hand valve 11 Re plug the power plug then start up the water purifier The indicators PO...

Страница 27: ...rtridge Sediment Filter 1 µm 10 included in kit ref 950019 8 Check that the O ring is properly positioned in the filter holder groove then re screw it A good tightening by hand is enough 9 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 10 Open the tap water hand valve 11 Re plug the power plug then start up the water purifier The indicat...

Страница 28: ...en remove with nippers the worn reverse osmosis membrane Introduce the new membrane the peripheral joint at the top up to complete block stop the extremity of the collector tube must outcrop the membrane holder one 10 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 11 Open the tap water hand valve 12 Re plug the power plug then start up t...

Страница 29: ... at the bottom of the water purifier in order to detect a possible leak during the starting up 8 Open the tap water hand valve 9 Re plug the power plug then start up the water purifier The indicators power pump and flush light the water purifier starts 2 minutes of rinsing cycle of the osmosis membrane The rinsing is ended when the indicator FLUSH switch off then the water purifier is on productio...

Страница 30: ...ter flow 7 Repeat the procedure on the OUT 8 Dry the floor at the bottom of the water purifier in order to detect a possible leak during the starting up 9 Open the tap water hand valve 10 Re plug the power plug then start up the water purifier The indicators POWER PUMP and FLUSH light the water purifier starts 2 minutes of rinsing cycle of the osmosis membrane The rinsing is ended when the indicat...

Страница 31: ...for the VERSION1 Ref 950242 Pressostat high pressure Ref 950225 Booster Pump 300GPD Ref 951301 Pressure gauge Ref 950231 Pressurised storage tank 75 Liters Ref 950240 Pressostat low pressure Ref 950290 VERSION 1 Controller Ref 950227 VERSION 2 Controller Ref 959050 Electrical box ...

Страница 32: ...Version 2020 02 10 31 Solenoid valve 3 ways Ref 951115 Conductimeter sensor Ref 951107 Pressure sensor 3 Bar Ref 951114 Solenoid valve for the VERSION 2 Ref 950277 Power supply Ref 950520 Filter holder 10 white body Ref 950238 ...

Страница 33: ...d analyser Purified water exit resin bottle Hand valve By pass Reverse Osmosis Membrane Flow restrictor Hand valve of storage tank Solenoid valve Pump Hand valve Exit for testing reverse osmosis membrane Pressure gauge pump exit Solenoid valve High pressure switch Low pressure switch Pressure gauge tank exit 5 micron sediments filter Carbon block cartridge 1 micron sediments cartridge Check valve ...

Страница 34: ... valve of storage tank Solenoid valve Pump Hand valve Exit for testing reverse osmosis membrane Pressure gauge pump exit Solenoid valves Pressure gauge tank exit 5 micron sediments filter Carbon block cartridge 1 micron sediments cartridge Check valve To drain Low pressure switch Inlet water The hydraulic wiring Tubing 3 8 Tubing 1 4 Check valve Pressurised storage tank osmosed water 3 ways valve ...

Страница 35: ... VALVE 1 4 Reference 950330 Quantity 4 SYSTEM FOR MEMBRANE TEST PEN CONDUCTIVITY METER TEE UNION 1 4 Reference 950091 Quantity 2 TWO WAY DIVIDER 1 4 Reference 950094 MALE CONNECTOR 1 4 1 2 NPTF Reference 950084 Quantity 2 TUBE ELBOW UNION 1 4 Reference 950089 Quantity 4 HOUSING 20 FILTER KEY HOUSING 10 FILTER KEY HOUSING MEMBRANE KEY1 HOUSING MEMBRANE KEY2 Drain clamp Water inlet valve POLYETHYLEN...

Страница 36: ...020 02 10 35 MANUFACTURER DiaSys Technologies 1682 rue de la Valsière Cap Gamma Parc Euromédecine II 34790 GRABELS Tél 33 0 4 11 95 03 40 Fax 33 0 4 11 95 03 50 Internet http www diasys diagnostics com Email info diasys technologies com ...

Отзывы: