O
perator’s manual
version: 2021.06
31
Change of the 2 RESIN CARTRIDGES
1.
Close the tap water hand valve.
2.
Close the storage tank hand valve.
3.
Open the water purifier exit hand valve to reduce the water pressure inside the water purifier; on
the controller screen is displayed «NO FEED». Keep the water purifier exit hand valve open.
4.
Turn off the water purifier power supply (power switch on position « 0 »), then unplug the power
plug.
5.
Remove the two transparent covers.
6.
Using the big key (supplied with water purifier) carefully unscrew (turn to the left) the white filter
holder (the left in front);
WARNING: the filter is filled with water, a mop is recommended!
7.
Remove the filter cartridge. Notice the orientation of the used cartridge. Introduce the new 0.75L
resin cartridge
(
included in kit ref. 959030)
8.
Check that the O-ring is properly positioned in the filter holder groove, then re-screw this one. A
« good tightening by hand » is enough.
TAKE CARE TO THE DIRECTION !
Up
Bottom
Remove the 0.75L
resin cartridge from its
holder
0.75L resin cartridge housing