1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 26 of 58
1.3.1.7. Electrode Tray Assembly
This assembly is comprised of the electrode tray and the sample sensor, AV-BP5036D. (
Figure 1-31 Electrode Tray with Sample Sensor
To remove the electrode tray assembly from the analyzer;
1. First, if the instrument is on, place in Hold in Maintenance mode to prevent reagents from
cycling. (
Main Operations Menu>>>Maintenance>>>Hold in Maintenance
)
2. Disconnect the sample sensor cable from the analyzer. (
3. Remove the sample intake tubing on the right side of the assembly where the sample sensor
is located. (
Figure 1-32 Electrode Tray with electrodes installed
4. Unplug the Reference Tubing Connector
5. Disconnect the waste tube (green tag) from the left side of the electrode tray.
6. Use one hand to press the Electrode Tray Plastic Tabs toward the center.
Sample Sensor
Electrode Tray
Base
Sample Intake
Tubing
Sample Sensor
Connector
Reference Tubing
Connector
Waste Tube
(tagged Green)