![Diamond SmartLyte Plus Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/diamond/smartlyte-plus/smartlyte-plus_service-manual_2493456015.webp)
1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 15 of 58
6. Disconnect the communication ribbon cable from the External Communication Board on the
back panel leading to the Display Board.
7. Disconnect the power supply cable attached to the CPU board by depressing the locking clip
on the left side of the connector and gently pulling the power cable out.
8. On the left side of the chassis, pull the top red grounding wire connector off.
9. Separate the rear panel from the front panel.
1.3.1.3. Power Supply Subassembly
To access the Power Supply Subassembly, the Rear Panel must be removed first.
1. Ensure the power is off.
2. Remove the Rear Panel according to
3. Disconnect the Power Inlet Module Connector (
Figure 1-10 Rear Panel with Power Supply and Main Power Receptacle
4. Remove Ground Cables from Power Inlet Module and left side of Dual Tab.
5. Finally, remove the four screws securing Blue Ring Connector and the Power Supply
Mounting Plate to the Rear Panel. (
6. The Power Supply Subassembly can be removed, (
Figure 1-11 Power Supply and Ground Cable Assemblies
Power Supply
Connector to CPU
Board
Power Supply
Connect To
Mounting Plate
Black Power Inlet
Module
Red Ground
Connectors
Power Inlet
Module Connector
Содержание SmartLyte Plus