Method for disposing of the device 

If  the  product  is  returned  to  the  manufacturer  at  the  end  of  its  life  the  company  will  ensure 

disposal in line with the Waste Electrical and Electronic Equipment Directive (WEEE) 2012/19/EU 
 
 
 

 
 
 
 

 
 
 

An additional water trap may be 
added to remove condensed 

water from the system 

Содержание Baby CPAP

Страница 1: ...P INSTRUCTIONS FOR USE MANUAL Diamedica UK Ltd Grange Hill Industrial Estate Bratton Fleming Barnstaple Devon EX31 4UH United Kingdom Tel 44 0 1598 710066 Fax 44 0 1598 710055 Email support diamedica co uk www diamedica co uk Revision B 31 05 2018 DCN 0012 ENG 0088 ...

Страница 2: ...oncentrator is also included and must be reviewed for specific safety and maintenance requirements before use The policy of Diamedica UK Ltd is to continuously improve its products Changes may be made to this manual without notice being given Users of the Diamedica Baby CPAP must read understand and follow the guidance given in this manual before using the system THE NEED FOR PATIENT MONITORING WA...

Страница 3: ...oncentrator delivers both oxygen and continuous positive airway pressure the level of which depends on the distance the tubing is under the water in the bottle Main gas outlet Expiratory limb Humidifier bottle Bottle with dial up CPAP level Air flow meter Oxygen flow meter Hour meter Oxygen Air mixing chart Start button Inspiratory limb Only persons who have read and understood this entire manual ...

Страница 4: ... yellow light should turn on and remain lit for a few minutes usually within 10 minutes until the concentration of the gas supplied reaches and exceeds 85 3 O The yellow light should then turn off after this time showing that the concentrator is running correctly 4 Fill the bottle to the level indicated with boiled water that has been allowed to cool and screw the lid back on 5 Connect the inspira...

Страница 5: ... of air and oxygen required to see a constant stream of bubbles in the water Start with 5 L min of oxygen and 5 L min of air Check and record the flows dialled and the oxygen concentration this provides A total flow of 10 L min is sufficient in most cases to deliver CPAP which you can see by continuous bubbling in the bottle 10 Check the child for signs of respiratory distress check the SpO2 and c...

Страница 6: ...st done first thing in the morning when there is likely to be adequate staff to observe the child throughout the day If trials off CPAP are started in the late afternoon low staff numbers overnight and the oxygen desaturation that sometimes occurs during sleep mean that there is a risk of hypoxaemia developing unrecognized overnight Children who have an SpO2 90 while still on CPAP or who are unsta...

Страница 7: ... essential to replace the one that is being cleaned without interruption to the child s treatment The CPAP circuit inspiratory and expiratory limb bottle and lid and connections must be thoroughly cleaned as follows 1 Wipe off any gross soiling Clean first with detergent soap to remove gross contamination e g blood sputum 2 Rinse with water that has been boiled and allowed to cool to tepid Let it ...

Страница 8: ...nd bottle in a clean plastic bag labeled and dated Store in the dry and clean area separate from a soiled equipment area Routine Cleaning and Maintenance of the CPAP Concentrator Each week the oxygen concentrator will require approximately 30 minutes of attention Concentrators have a large particulate filter over the air inlet opening usually at the back of freestanding or portable models This fil...

Страница 9: ...device If the product is returned to the manufacturer at the end of its life the company will ensure disposal in line with the Waste Electrical and Electronic Equipment Directive WEEE 2012 19 EU An additional water trap may be added to remove condensed water from the system ...

Страница 10: ...atic Voltage Switcher rated at 30 Amps The AVS30 will switch off the equipment connected to it if the mains power goes outside pre set acceptable limits and will reconnect automatically when the mains power returns to normal Reconnection takes place after a delay to ensure stability of the mains OR OR Input good GREEN LIGHT Power OK Input unstable RED LIGHT No Power Input now good AMBER LIGHT No P...

Страница 11: ...1 of 13 Copyright Diamedica UK Ltd 2018 DIAMEDICA UK LTD Grange Hill Industrial Estate Bratton Fleming Barnstaple Devon EX31 4UH UK Tel 44 0 1598 710066 Email support diamedica co uk www diamedica co uk 0088 ...

Страница 12: ...of manufacture Indicates the date when the medical device was manufactured Use by date Indicates the date after which the medical device is not to be used Batch code This symbol shall be accompanied by the manufacturer s batch code The batch code shall be adjacent to the symbol Catalogue number Indicates the manufacturer s catalogue number so that the medical device can be identified Serial number...

Страница 13: ...mportant cautionary information such as warnings and precautions that cannot for a variety of reasons be presented on the medical device itself Non sterile Indicates a medical device that has not been subjected to a sterilization process Class II equipment Type BF applied part Recycling symbol Products with this symbol should not be disposed of in the bin The battery recycling symbol Chemical symb...