Diagnosis Optilet Mini Скачать руководство пользователя страница 2

REMOVING THE USED LANCET

CLEANING AND DISINFECTION OF THE PRICKER

PRICKING IN THE ALTERNATIVE SITE

1. 

Remove the pricker adapter

Remove carefully the adapter from the 
pricker.

 

Be careful not to touch the 

uncovered lancet.

1. 

Place the lancet

Place the lancet in lancet carrier; press it to 
the end inside the appliance.

2. 

Install the transparent adapter for the 

alternative site testing (AST).

Install the transparent adapter for the 
alternative site testing (AST) in place of the 
pricker adapter.
The adjustment of the puncture depth is not 
required if the (AST) adapter is used.

3. 

Pull the sleeve

 

Pull the slide-on sleeve forward until the 
click is heard.

4. 

Hold down the transparent adapter on 

the skin

Press down the pricker stably and then 
press the release button until the sufficient 
blood volume appears in the adapter.

5. 

Lift up carefully the pricker vertically.

Be careful not to spread the blood sample at 
the puncture site. Apply a drop of blood on 
the test strip.

2. 

Put the cover on the lancet

Before the removal of the lancet, put the 
protective lancet cover on the hard surface, 
then stab carefully the lancet in it. 

3. 

Remove the lancet

Slide the lancet ejector forward until the 
lancet pulls out of the pricker. Remove the 
used lancet.

4. 

Install the adapter

Reinstall the pricker adapter on the 
appliance.

IMPORTANT: 

If you consider the use of the other puncture 

site than the fingertip, 

consult a physician.

The transparent adapter for pricking in the alternative sites 

(forearm or palm of the hand) is designed for the use with 

Optilet Mini in order to collect a blood sample for measuring 

the concentration of glucose.

For use only by a single patient.

Read the instructions for use of glucose meter in relation to the 

collection of blood samples from the alternative sites.

In order to clean the pricker, wipe the elements with a soft cloth wet 
with water and mild detergent. Do not submerge the appliance in any 
liquid. In order to disinfect the elements, prepare a solution of 
household bleach and water in the ratio 1:9. Wipe the appliance with 
a soft cloth wet with this solution. Submerge only the adapter in this 
solution for 30 minutes. After disinfection, rinse and let it dry in the 
air.

FUNCTION OF PRECISE CLEANING

If blood enters the interior of the pricker adapter, turn the 
transparent pricker tip to the position "

" (Fig. A);

Remove the transparent pricker tip (Fig. B) and wipe it with a soft 
cloth wet with the afore-mentioned solution for disinfection (Fig. C).

REMEMBER

:  

While reinstalling the transparent pricker tip, check if 

the depth indicator is set in the position "

" (Fig. A), then set the 

puncture depth appropriate for yourself.

Transparent adapter for 
alternative site testing 
(AST)

Release button

Lancet 

ejector

Slide-on 

sleeve

Expiry date

Read the instructions for use

Producer

Date of last revision

Not intended for re-sterilisation

Catalogue number 

Lot number

Sterile product

For single use

Number of pieces in set

Do not use if the package is damaged

Warning

Date of production

Diagnosis S.A.
ul. Gen. W. Andersa 38A
15-113 Białystok, POLAND

Marking for OPTILET mini

Marking for OPTILETS

5221

0197

Rev. 2018.01.03 /EN

A

B

C