Di-LOG DL6702 Скачать руководство пользователя страница 2

III.  Specifications: 

Operating environment:

 0 oC to 50 oC (32 oF to 122 oF) at < 70 % relative humidity. 

Storage temperature:

 -20 oC to 60 oC (-4 oF to 140 oF) at < 80 % relative humidity. 

Power:

 One 9V, NEDA1604, IEC6F22 Battery. 

 

Dimensions:

 209 x 43 x 30mm. 

Weight: Approx.: 

270g. 

 

Accuracy is given at (22 oC~ 28 oC) / (71 oF~ 82 oF), less than 70%RH 

 

Function 

Range 

Output 

Sensitivity   

Accuracy %rdg 

Overload Protection 
(No input) 

200A 

1mV/A 

1A/mV DC 

+(3.0% + 0.6Amps) 

MAX. 200A Output 

DC Current   

600A 

0.1mV/A 

10A/mV DC 

+(3.5% +3Amps) 

MAX. 600A Output 

200A 

1mV/A 

1A/mV AC 

+(3.5% + 0.4Amps) 

MAX.200A Output 

AC Current 
(Frequency : 
50Hz~400Hz) 

600A 

0.1mV/A 

10A/mV AC 

+(4.0% + 4Amps) 

MAX. 600A Output 

 

IV. Operation Instructions 

 

DC Current Measurements: 

a.  Set the rotary switch at 200A DC or 600A DC. 
b.  Push  the  Zero  Button  to  set  the  reading  zero.  If  cannot,  release  the  button  a  while  and  push  it 

again. 

c.  Press  the  trigger  to  open  the  jaw  and  fully  enclose  the  conductor  to  be  measured.  No  air  gap  is 

allowed between the two half jaws. 

d.  Connect the test leads to the output shock of the clamp probe and input shock of the DMM. 
e.  Turns on the DMM and set it at DCV 400mV or 2V range. 
f. 

Read the measured value from the DMM LCD display. 

g.  Make sure that the offset value caused by the residual magnetism is still removed. If the new offset 

value is produced, remove it with the zero button and make a new measurement again according to 
the “c” and “d” (If the current to be measured is larger than the current to be measured before, or 
the direction of current changes, the new offset value will be produced) 

 

AC Current Measurements: 

a.  Set the rotary switch at200A AC or 600A AC. 
b.  Press  the  trigger  to  open  the  jaw  and  fully  enclose  the  conductor  to  be  measured.  No  air  gap  is 

allowed between the two half jaws. 

c.  Connect the test leads to the output shock of the clamp probe and input shock of the DMM. 
d.  Turns on the DMM and set it at ACV 400mV or 2V range. 
e.  Read the measured value from the DMM LCD display. 

 

V

Battery Replacement 

 

When the low battery LED lamp is turned on, replace the old battery with new one. 
a.  Turn the power off and remove the test leads from the clamp mater. 
b.  Remove the screw of the battery compartment. 
c.  Slide off the old battery. 
d.  Remove the old battery 
e.  Insert one 9V NEDA1604, IEC6F22 battery. 
f. 

Replace the battery compartment and secure the screw. 