III. Specifications:
Operating environment:
0 oC to 50 oC (32 oF to 122 oF) at < 70 % relative humidity.
Storage temperature:
-20 oC to 60 oC (-4 oF to 140 oF) at < 80 % relative humidity.
Power:
One 9V, NEDA1604, IEC6F22 Battery.
Dimensions:
209 x 43 x 30mm.
Weight: Approx.:
270g.
Accuracy is given at (22 oC~ 28 oC) / (71 oF~ 82 oF), less than 70%RH
Function
Range
Output
Sensitivity
Accuracy %rdg
Overload Protection
(No input)
200A
1mV/A
1A/mV DC
+(3.0% + 0.6Amps)
MAX. 200A Output
DC Current
600A
0.1mV/A
10A/mV DC
+(3.5% +3Amps)
MAX. 600A Output
200A
1mV/A
1A/mV AC
+(3.5% + 0.4Amps)
MAX.200A Output
AC Current
(Frequency :
50Hz~400Hz)
600A
0.1mV/A
10A/mV AC
+(4.0% + 4Amps)
MAX. 600A Output
IV. Operation Instructions
DC Current Measurements:
a. Set the rotary switch at 200A DC or 600A DC.
b. Push the Zero Button to set the reading zero. If cannot, release the button a while and push it
again.
c. Press the trigger to open the jaw and fully enclose the conductor to be measured. No air gap is
allowed between the two half jaws.
d. Connect the test leads to the output shock of the clamp probe and input shock of the DMM.
e. Turns on the DMM and set it at DCV 400mV or 2V range.
f.
Read the measured value from the DMM LCD display.
g. Make sure that the offset value caused by the residual magnetism is still removed. If the new offset
value is produced, remove it with the zero button and make a new measurement again according to
the “c” and “d” (If the current to be measured is larger than the current to be measured before, or
the direction of current changes, the new offset value will be produced)
AC Current Measurements:
a. Set the rotary switch at200A AC or 600A AC.
b. Press the trigger to open the jaw and fully enclose the conductor to be measured. No air gap is
allowed between the two half jaws.
c. Connect the test leads to the output shock of the clamp probe and input shock of the DMM.
d. Turns on the DMM and set it at ACV 400mV or 2V range.
e. Read the measured value from the DMM LCD display.
V
.
Battery Replacement
When the low battery LED lamp is turned on, replace the old battery with new one.
a. Turn the power off and remove the test leads from the clamp mater.
b. Remove the screw of the battery compartment.
c. Slide off the old battery.
d. Remove the old battery
e. Insert one 9V NEDA1604, IEC6F22 battery.
f.
Replace the battery compartment and secure the screw.