![Di-LOG 9083P Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/di-log/9083p/9083p_operating-instructions-manual_2492505013.webp)
9083P
Multifunction Tester
Operating Instructions
328A561 revision1.1
- 13 -
4.2 LCD display
1. Mains supply status icons
These icons indicate the status of the mains supply
between phase-earth (PE), phase-neutral (PN) and
phase-earth (PE) during RCD and Loop tests.
Note: Testing is inhibited if the mains supply is
incorrect.
2. Analog bargraph
The analogue bargraph provides an analogue
representation of the measurement shown by the
primary digital display.
3. Primary digital display
The primary digital display is used to indicate the test
measurement for the active test type.
4. Icons for function key F1.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F1
cycles through the available options.
5. Icons for function key F2.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F2
cycles through the available options.
6. Icons for function key F3.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F3
cycles through the available options.
7. Secondary display for function key F4.
The secondary display is used to show the measured
output
voltage
during
insulation
resistance