MFC SWITCHBOX USER’S MANUAL
© 1997-2002 DH Instruments, Inc.
Page 10
5.4
EXAMPLE .CFG FILE
This file keeps a history of the current and previous analog calibration values. The name of the file is
based on the molbox sn #.
The UPDATE line represents the last time the particular value was updated.
The order of the values is: set adder, set mult, sense adder, sense mult, measure adder, and measure mult.
This file should not exceed 25 lines including help text, nor should there be any other colons other than
those below, Analog uses them as a delimiter. The order and spacing of the following lines should also
be the same.
Current Update:
Dec. 28, 1995 at 7:44:34 AM
Current Value:
0.001, 0.970246, 0.002, 1.019117, 0.004, 1.015546
Previous Update:
Nov. 28, 1995 at 7:41:44 AM
Previous Value:
0.001, 0.971293, 0.002, 1.018018, 0.005, 1.013785
Former Update:
Oct. 28, 1995 at 7:36:41 AM
Former Values:
0.002, 0.973394, 0.001, 1.016267, 0.003, 1.011819
Initial Creation:
Sept. 28, 1995 at 7:33:18 AM (file creation date)
Initial Values:
0.002, 1.023425, 0.003, 0.965949, 0.005, 0.961934