GB-K Gas Booster Package Operation and Maintenance Manual
Page 19
© 1998 - 2000
DH Instruments, Inc
.
3.4.3
DECREASING PRESSURE
The High Pressure Outlet Regulator (7) is a venting type. Therefore, it is possible to regulate
pressure down without gas flow through the regulator.
n
Adjust the pressure down using the High Pressure Outlet Regulator (7) by rotating the
knob counter-clockwise until a pressure below the setpoint is achieved as indicated on
the High Pressure Outlet Gauge (4).
o
Set the pressure to the final setpoint by rotating the knob clockwise until the desired
pressure is indicated on the high pressure gauge.
Содержание GB-K-152
Страница 12: ...GB K Gas Booster Package Operation and Maintenance Manual 1998 2000 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 18: ...GB K Gas Booster Package Operation and Maintenance Manual 1998 2000 DH Instruments Inc Page 12 N NO OT TE ES S...
Страница 26: ...GB K Gas Booster Package Operation and Maintenance Manual 1998 2000 DH Instruments Inc Page 20 N NO OT TE ES S...
Страница 28: ...GB K Gas Booster Package Operation and Maintenance Manual 1998 2000 DH Instruments Inc Page 22 N NO OT TE ES S...
Страница 34: ...GB K Gas Booster Package Operation and Maintenance Manual 1998 2000 DH Instruments Inc Page 28 N NO OT TE ES S...