DC-3000 SERIES ULTRASONIC THICKNESS GAUGE
- 10 -
error. If the error is not enough to influence the measuring accuracy, it can
be neglect; otherwise it is necessary to get the accurate sound velocity of
the workpiece to be measured. The
“velocity measurement” can be used,
the usage is as follows:
Select T-E mode, Velocity can be neglect, and directly measure the sample
which thickness is known.
Press
key into the menu-
“SETTING”
-
“Vel. Rate”
-
“Vel
measurement”,
Press
or
to move the arrow, press
or
to up and
down the value of velocity to determine the thickness as the same as the
value of sample that is measured,
Press
to store this new velocity,
Press
or
to select location where you want to store,
Press
to confirm.
8.5.1.4 Velocity storage
This function allows users to store 4 new velocities Rate.
8.5.2
Resolution
Users can select the displayed resolution. When selecting the high
accuracy, the workpiece surface to be measured should be smooth, for
the purpose of getting an accurate value.
Press
key into “Resolution”,
Press
or
to set resolution and unit.