DC-2000C SERIES ULTRASONIC THICKNESS GAUGE
11
velocity, press
to select one, press
to confirm.
- Press
to Esc. Menu and into the measurment.
This new velocity will be stored. And it can be found from
“2. Velocity
rate”- “
⑷
Vel. Storage”
for further use.
8.4.2.4 Velocity Storage
DC-2000 series provides 4 locations to store new velocities.
〖
8.4.3
〗
Resolution
- Press
key into “Resolution”
- Press
or
to set resolution and unit.
1. 0.1 mm
2. 0.01 mm
3. 0.01 in
4. 0.001 in
- Press
key to enter/confirm
〖
8.4.4
〗
Probe Calibration
It will cause error during the primary stage of usage and operating. If this
caused by the probe itself, please use following calibration method:
- Measure the test piece with known thickness.
- Press
into
”4.Probe calibration”
- Press
or
into “Calibration”