 INDICATIONS FOR USE

User's Manual | 

MDP156

X

Recommended separation distances between portable and mobile RF communications 

equipment and the product

The product is intended for use in an electromagnetic environment in which radiated RF distur-

bances are controlled. The customer or the user of the product can help prevent electromagnetic 

interference by maintaining a minimum distance between portable and mobile RF communica-

tions equipment (transmitters) and the product as recommended below, per the maximum out-

put power of the communications equipment.

Rated maximum output 

power [W] of transmitter

Separation distance per frequency of transmitter [m]

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 1.2

800 MHz to 2.5 GHz

d = 2.3

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated a maximum output power not listed above, the recommended separa-

tion distance d in meters (m) can be estimated using the equation applicable to the frequency 

of the transmitter, where P is the maximum output power rating of the transmitter in watts 

(W) per the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range 

applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is af-

fected by absorption and reflection from structures, objects and people.

X

General Safety

product is a DC powered device. Use with the supplied AC adaptor. Operate the unit on 

100-240V AC only.

For plug-in equipment, the power outlet socket must be located near the equipment and 

must be easily accessible.

Use a power cord that matches the voltage of the power outlet, which has been approved 

and complies with the safety standard of your particular country.

The nameplate indicating operating voltage, etc. is located on the AC adaptor.

Should any solid object or liquid fall into the cabinet, unplug the unit and have it checked 

by qualified personnel before operating it any further.

Unplug the unit from the wall outlet if it is not to be used for several days or more.

To disconnect the AC power cord, pull it out by grasping the plug. Never pull the cord it-

self.

The socket-outlet shall be installed near the equipment and shall be easily accessible.

X

Healthcare use

Please do not touch patient and this healthcare device at the same time.

Users must not allow SIP/ SOPs and the patient to come into contact at the same time.

Grounding reliability can only be achieved when the equipment is connected to an equiva-

lent receptacle marked “Hospital Only” or “Hospital Grade”.

X

Installation

This device can only be installed by manufacturer authorized and trained personnel.

Prevent internal heat build-up allowing adequate air circulation.

Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draper-

ies) that may block the ventilation holes.

Do not install the unit near heat sources such as radiators or air ducts, or in a place sub-

ject to direct sunlight, excessive dust, mechanical vibration or shock.

Do not place the monitor near equipment which generates magnetism, such as a trans-

former or high voltage power lines.

Use suitable mounting apparatus to avoid risk of injury.

X

Precautions for connecting this unit with other healthcare devices

Before you utilize this device and/or connect this device to any other healthcare device, please 

be aware of and abide by the following precautions:

Before actually using this device for healthcare practice, please check and confirm that 

you do not experience any discomfort in the use of this monitor

If you experience or are likely to experience discomfort, please refrain from using this de-

vice.

Generally, discomfort (such as eye strain, fatigue, nausea, or motion sickness) can be pro-

voked by quick movements of video picture, focal positioning of video images, distances 

between moving objects and changing image colors.

Before prolonged use, make sure the image of the connected healthcare device is dis-

played properly.

X

Use with an electrosurgical knife, etc.

If this unit is used together with an electrosurgical knife, etc., the picture may be disturbed, 

warped or otherwise abnormal because of strong radio waves or voltages from the device. This 

is not a malfunction. When you use this unit simultaneously with a device from which strong 

radio waves or voltages are emitted, confirm the effect of this before using such devices, and 

install this unit in a way that minimizes the effect of radio wave interference.

X

Precautions for using this unit safely.

Some people may experience discomfort (such as eye strain, fatigue, or nausea) while 

watching video images. It is recommended that all viewers take regular breaks while 

watching video images. The length and frequency of necessary breaks will vary from per-

son to person. You must decide what works best.

Avoid watching the display in environments where your head may shake, because there is 

a higher possibility that you experience discomfort.

Содержание MDP156

Страница 1: ...MDP156 15 6 Healthcare Display User s Manual A 605 M 2034...

Страница 2: ...CC and DOC Statement on Class B This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC rules These limits are designed to provid...

Страница 3: ...notices for use in the healthcare applications 11 Guidance and manufacturer s declaration electromagnetic emissions 12 Guidance and manufacturer s declaration electromagnetic immunity 12 Recommended s...

Страница 4: ...devices even before installing them in your system unit Static electrical discharge can damage computer components without causing any signs of physical damage You must take extra care in han dling t...

Страница 5: ...dapter Power Cord VGA Cable HDMI Cable USB Cable for Touch Optional Items Cable Clamp x 3 pcs VESA Desk Stand The main body and accessories in the package may not come similar to the information liste...

Страница 6: ...operation Mechanical Specification Mounting VESA Mount 75 x 75 mm 100 x 100 mm 200 x 100 mm VESA Screws M4x4L 75x75mm 100 x 100mm flat head Dimensions W x H x D 387 8 x 232 9 x 38 45 mm Housing Back c...

Страница 7: ...to be played through the monitor speak ers 4 HDMI Transmits and protects copyrighted digital video and audio Example An HD tuner to an HD ready TV 5 VGA RGB Transmits video from a PC to a monitor Exam...

Страница 8: ...ews to fix the desk stand to VESA holes on the back cover of the device Important The picture is for demonstration purposes only VESA Mount accessories are not supplied by default Connecting Periphera...

Страница 9: ...To turn off the device please disconnect the power cord from an electric outlet Important To fully disengage the power please disconnect the input plug from the ac outlet To avoid risk of electric sh...

Страница 10: ...cles trains high volt age lines or fluorescent lamps Bad color Color density or tint control may not be adjusted properly See the adjusted value of image Image appears and disappears Input may not be...

Страница 11: ...n shall be addressed to the authorized representative For any service or guarantee matters please refer to the ad dresses provided in the separate service or guarantee documents X Important safeguards...

Страница 12: ...c dis charge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ce ramic tile If floors are covered with synthetic material the relative humidity should...

Страница 13: ...by an electromagnetic site survey A should be less than the compli ance level in each frequency range B Interference may occur near equipment marked with following symbol NOTE 1 At 80 MHz and 800 MHz...

Страница 14: ...the patient to come into contact at the same time Grounding reliability can only be achieved when the equipment is connected to an equiva lent receptacle marked Hospital Only or Hospital Grade X Insta...

Страница 15: ...ages for extended periods or using the unit repeatedly in a high temperature high humidity environments may cause image smearing burn in areas of which brightness is permanently changed lines or a dec...

Страница 16: ...each single failed Product that is returned for servicing repair reworking or failure investigation to Seller or to the Authorized Service Provider At least one adhesive yellow label shall be attached...

Отзывы: