background image

 

Programs C, D & E

 

 

Procedure  for  Testing  Low  Levels  of  Water  in  Oil  Using  a  5  mL 
Sample Size 

 

(Reagent System: HS-LRP) 

 
 
1.

 

PREPARATION:

  Remove a sampling syringe and a white-capped reaction tube from the reagent box.  Check 

the contents to ensure that all items are present and intact.   Remove the  white cap from  the reaction tube and 
discard.  Carefully remove a green water ampule from the cardboard carton and insert it fully into the top of the 
protective ampule  sleeve containing the  grayish  white ampule. 

Do  not replace the  protective ampule  sleeve 

into the tube at this time.

 

 

2.

 

SAMPLING:

  Using a clean polyethylene sampling pipette, pipette 5 mL of the oil to be tested into the reaction 

tube, e.g., up to 5 mL line marked on side of the reaction tube 

(NOTE: The protective ampule sleeve is not to 

be in the tube at this time.)

.  Insert the protective ampule sleeve into the reaction tube with the grayish white 

ampule  on  the  bottom,  then  insert  the  black  septum  (hole  on  top)  into  the  reaction  tube  by  gripping  the  tube 
firmly and pressing down on the septum with your thumb until flush with top of the tube.  

TIGHTLY

 cap the 

tube with the holed, white cap. 

NOTE:

 

A homogeneous, representative sample of the oil being tested must 

be  used  to  obtain  accurate  results. 

  DO  NOT  SQUEEZE  THE  SIDES  OF  THE  TUBE  WHILE 

INSERTING THE BLACK SEPTUM.  DOING THIS WILL RESULT IN LOW READINGS.

 

 

3.

 

REACTION:

    Break  the  bottom  (grayish  white)  ampule  in  the  reaction  tube  by  compressing  the  sides  of  the 

tube.  Mix thoroughly by shaking the reaction tube vigorously for 30 seconds.  Allow the reaction to proceed for 
a  total  of  4  minutes.  The  mixture  will  bubble  and  perhaps  foam  during  this  time.   

DO  NOT  BREAK  THE 

GREEN AMPULE.

  (NOTE: If the test sample is very thick and/or viscous, shake the reaction tube again for 

30 seconds and let stand for an additional 4 minutes.) 

 
4.

 

ANALYSIS:

  Press the <ON> key once to activate the meter.  Ensure the meter is set for the correct program C, 

D or E for the sample matrix being tested (See HydroSCOUT Manual on “Selecting a Program” for instructions 
on how to change programs.).  Once the meter displays 

tUbE

”, insert the reaction tube into the opening at the 

bottom of the meter.  The hole in the white cap must be lined up with the metal pin extending from the inside of 
the hole of the meter.  Push the reaction tube all the way into the meter.  Press the <READ> key on the face of 
the  meter.  The display will flash the  word “

CALC

,” and the results will be displayed 

micrograms  of  water 

per milliliter of sample.

  Record the results.  

KEEP THE METER AND REACTION TUBE UPRIGHT AT 

ALL TIMES.

 

 
5.

 

DISPOSAL:

  Carefully  remove  the  reaction  tube  from  the  meter.    Vent  the  reaction  tube  by  inserting  it 

completely into the specially designed venting cap unit and remove. 

REACTION TUBE AND CAP MUST BE 

UPRIGHT  AND  POINTED  AWAY  FROM  USER  AND  BYSTANDERS.

    Tap  the  reaction  tube  on  a  hard 

surface to shake any loose glass to the bottom of the tube, then break the top (green) ampule in the reaction tube 
by compressing the sides of the reaction tube.  Shake the tube for 15 seconds.  After 5 minutes, vent the tube 
again.  Discard the reaction tube as laboratory waste. 

DO NOT REMOVE WHITE CAP.

 

6.

 

CALCULATION:

  The meter is programmed to display the result in 

micrograms per milliliter

 

(µg/mL)

.  If a 

result is to be reported as a weight/weight value, divide the HydroSCOUT result by the density of the oil. 

 
 

 

Water in Oil (µg/g) = R/d 

 
 

where: 

 
 

 

R = Meter reading in µg/mL 

 

 

d = Density of oil in g/mL 

 

Содержание HydroSCOUT

Страница 1: ...aims any implied warranties of merchantability or fitness for any particular purpose Further Dexsil Corporation reserves the right to revise this publication and to make changes from time to time in t...

Страница 2: ...BROKEN GLASS AS IT CAN PENETRATE THE PLASTIC TUBE AND CUT FINGERS THE GRAY AMPULE IN THE TUBES CONTAINS CALCIUM HYDRIDE WHICH IS A FLAMMABLE SOLID AND IS WATER REACTIVE IN CASE OF ACCIDENTAL BREAKAGE...

Страница 3: ...gram A Procedure for Testing Water Content up to 20 6 7 Program B Procedure for Testing Water Content up to 100 7 Program C D E Procedure for Low Levels of Water 5 mL Sample Size 8 Program F Procedure...

Страница 4: ...ogram is described below Table 1 HydroSCOUT Analysis Program Information Program Sample Matrix Types Tested1 Units MDL3 Maximum Reagents Needed Sample Size A Used Oil Liquid Used Oil Chemtool 250 Pain...

Страница 5: ...the pressure function is a linear parameter the meter response value can be corrected to produce accurate results on all oil types The HydroSCOUT response on low water level oils has been documented...

Страница 6: ...thod D 7358 In order to achieve a result based on the weight of the sample the sample must be weighed into the reaction vessel Not only does this allow for a weight based result but it improves accura...

Страница 7: ...to G followed by OFF and then back to A 0 20 again Press the READ key at any point to fix the selection at the displayed program or to turn the meter off when OFF is displayed To begin using the selec...

Страница 8: ...i e programs C D or E ethylene glycol and acids have been found to produce a positive result Ethylene glycol responds at approximately 7 of the volumetric equivalent Error Code Explanation Likely Caus...

Страница 9: ...ule from the cardboard carton and insert it fully into the top of the protective ampule sleeve containing the grayish white ampule Return the protective ampule sleeve to the reaction tube with the gra...

Страница 10: ...present depress the plunger again and draw a second sample slowly Remove the syringe from the oil sample and wipe any excess oil from the outside of the syringe with a tissue wipe be sure not to blot...

Страница 11: ...bubble and perhaps foam during this time DO NOT BREAK THE GREEN AMPULE NOTE If the test sample is very thick and or viscous shake the reaction tube again for 30 seconds and let stand for an additiona...

Страница 12: ...TUBE WHILE INSERTING THE BLACK SEPTUM DOING THIS WILL RESULT IN LOW READINGS 4 REACTION Break the bottom grayish white ampule in the reaction tube by compressing the sides of the tube Mix thoroughly...

Страница 13: ...3 REACTION Break the bottom gray ampule in the tube by compressing the sides of the tube Mix thoroughly by shaking the tube vigorously for 30 seconds The mixture will be seen to bubble and perhaps fo...

Страница 14: ...ight The difference in the two weights becomes the sample weight S2 to be used in equation 2 below Insert the black septum hole on top into the reaction tube by gripping the tube firmly and pressing d...

Страница 15: ...0 0 50 01 50 61 1 HydroSCOUT results converted to w using sample density Analysis of Used Oil Certified Reference Materials on Programs A and B2 CRM Certified Value wt HydroSCOUT1 wt Karl Fischer wt E...

Страница 16: ...Oil Light 26 53 08 236 202 007 448 410 043 838 778 026 Shell Turbo T32 23 34 019 220 223 04 433 403 01 825 800 038 Ideal Turbine Oil AW 32 39 43 072 210 208 013 400 435 005 759 842 013 Renolin Turbin...

Страница 17: ...0 0 2 051 87 101 038 175 166 057 349 374 009 696 712 027 1397 1528 015 Valvoline Non Detergent SAE 30 0 30 059 88 96 042 175 136 061 350 368 028 701 784 02 1401 1497 041 Oilzum Non Detergent SAE 50 0...

Страница 18: ...172 126 047 343 289 035 687 602 039 1374 1313 015 Quaker State SAE 30 0 7 018 88 105 06 176 183 104 352 333 044 703 597 04 1406 1491 018 Quaker State Proline 10W 40 Used 0 26 079 89 195 076 178 294 0...

Страница 19: ...0 18 68 0 0184 29 9 26 63 0 0081 Topsoil 0 0 21 5 0 5 13 0 0377 10 1 10 33 0 0248 20 0 19 50 0 0026 Sand 0 0 22 5 1 5 30 0 0266 10 1 10 30 0 0274 15 0 15 42 0 0304 Clay Loam I 0 0 5 0 4 47 0 0354 10 2...

Страница 20: ...luid D Phenolic resin containing approximately 10 unreacted Phenol Fluid E Pigmented polyester polyol resin system containing MEK Toluene Ethyl Acetate PM Acetate and Ethyl 3 Ethoxy Propionate Fluid F...

Страница 21: ...wt G 0 15 wt wt 100 wt wt Using a 10 gram sample Using a 1 gram sample Program Storage Non Volatile Display 4 Digit 0 35 inch LCD with Program Indicator Battery One 9V Lithium included Use only Alkali...

Отзывы: