Manufacturer’s Declaration

The intraoral camera emits electromagnetic energy at levels compliant 
to international standards; however, interference between the camera 
and other devices is possible if located in close proximity. The customer 
or the user of the camera can help prevent electromagnetic interference 
by locating sensitive devices away from the camera.

This camera, classified as Medical Electrical Equipment, requires spe-
cial precautions regarding EMC and must be installed and put into ser-
vice according to the EMC information provided in the accompanying 
product documentation. Portable and mobile RF communications 
equipment can effect Medical Electrical Equipment. This camera com-
plies with EMC requirements when used with the cables and accesso-
ries supplied with the product. The use of accessories and cables other 
than those sold by DEXIS, LLC and specified as replacement parts for 
internal components, may result in increased emissions or decreased 
immunity of this camera. This camera should not be used adjacent to or 
stacked with other equipment. If adjacent or stacked use is necessary, 
this camera should be observed to verify normal operation in the con-
figuration in which it will be used.

Guidance and manufacturer’s declaration - electromagnetic emissions 
for all  
EQUIPMENT and SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic emissions

This camera is intended for use in the electromagnetic environment specified 
below. The customer or the user of this camera should assure that it is used in such 
an environment.

Emissions test

Compliance

Electromagnetic environment - 
guidance

RF emissions 
CISPR 11

Group 1

This camera uses RF energy only for its 
internal function. Therefore, its RF 
emissions are very low and are not 
likely to cause any interference in 
nearby electronic equipment.

RF emissions 
CISPR 11

Class B

The camera is suitable for use in all 
establishments other than domestic, 
and may be used in domestic 
establishments and those directly 
connected to the public low-voltage 
power supply network that supplies 
buildings used for domestic purposes, 
provided the following warning is 
heeded: 
 

Warning: 

This camera is intended for use by 
healthcare professionals only. This 
camera may cause radio interference or 
may disrupt the operation of nearby 
equipment. It may be necessary to take 
mitigation measures, such as re-
orienting or relocating the camera or 
shielding the location.

Harmonic 
emissions 
IEC 61000-3-2

Class A

Voltage 
fluctuations/ 
flicker emissions 
IEC 61000-3-3

Complies

DEXcam4 HD EN.book  Page 10  Monday, January 16, 2017  2:12 PM

Содержание DEXcam4 HD

Страница 1: ...Intra oral Camera User Manual English...

Страница 2: ......

Страница 3: ...lts This camera is designed to be operated by healthcare professionals who are educated and competent in the techniques described in this manual Visual impairment is permissible so long as the user ca...

Страница 4: ...ase note that this manual assumes that new users possess basic computer skills and an understanding of the Windows operating system Absent this experience we highly recommend that you obtain these ski...

Страница 5: ...l lenses One touch Focus Adjustment 5 position zoom stepper switch offers a focus mechanism to view intra oral and extra oral images without a lens change The face icon captures smile arch and portrai...

Страница 6: ...ill not function properly Install the Intra oral Camera WARNING No modification of this device is allowed WARNING Loading incompatible software or viruses may result in loss of patient data WARNING Do...

Страница 7: ...th on the camera for each new patient CAUTION The sheath must be removed after each patient The intra oral camera must be cleaned and disinfected after each patient After cleaning the camera and handp...

Страница 8: ...lace the camera in the camera holster to turn the camera off Handling After extensive use the camera focusing motor may need to be reset only as needed 1 Press the button to move the LED indicator to...

Страница 9: ...understorms stop using the camera Failure to do so may cause damage to the product Do not disassemble drop or otherwise subject the camera to hard shocks It may cause electric shock short circuit or f...

Страница 10: ...scribed herein under control number 4005847 to be in compliance with the applicable regulations Intertek is a Nationally Recognized Testing Laboratory by the Occupational Safety and Health Administrat...

Страница 11: ...fications USB Compatibility USB 2 0 or later Electrical Ratings 5VDC 500mA Requirements for safety approvals This device must be connected to a personal computer that meets the standard specified in I...

Страница 12: ...ation in the con figuration in which it will be used Guidance and manufacturer s declaration electromagnetic emissions for all EQUIPMENT and SYSTEMS Guidance and manufacturer s declaration electromagn...

Страница 13: ...ronment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 0 5 1 kV differential mode 0 5 1 2 kV common mode Mains power quality should be that of a typical commercial or hospital environm...

Страница 14: ...eld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of...

Страница 15: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MH 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7...

Страница 16: ...ce described herein does not contain parts which can be serviced DEXIS Customer Care Center United States and Canada Tel 1 888 883 3947 Fax 1 888 833 3947 E mail support dexis com Outside North Americ...

Страница 17: ...Cam ra intra orale Manuel d utilisation Fran ais...

Страница 18: ......

Страница 19: ...re utilis e par des professionnels de la sant comp tents ayant suivi une formation sur les techniques d crites dans ce manuel Une d ficience visuelle de l utilisateur est acceptable dans la mesure o l...

Страница 20: ...s poss dent des connaissances de base en informatique et une compr hension du syst me d exploitation Windows Sans cette exp rience nous vous recommandons fortement d acqu rir ces comp tences par un co...

Страница 21: ...positions comprenant un dispositif de mise au point pour visualiser les images intra orales et extra orales sans changer la lentille L ic ne visage prend des clich s du sourire de l arcade et du visa...

Страница 22: ...e fonctionnera pas correctement Installer la cam ra intra orale AVERTISSEMENT Aucune modification de cet appareil n est autoris e AVERTISSEMENT Le chargement de tout logiciel incompatible ou d un viru...

Страница 23: ...veau patient MISE EN GARDE Il faut retirer la gaine apr s l examen de chaque patient La cam ra intra orale doit tre nettoy e et d sinfect e entre chaque patient Une fois que la cam ra et la pi ce main...

Страница 24: ...ment dans l tui et que son cordon est vers le bas La cam ra ne s teint pas si elle est plac e dans l tui avec son cordon vers le haut 5 Placer la cam ra dans l tui pour l teindre Manipulation Apr s un...

Страница 25: ...dommag e Des blessures pourraient en r sulter viter d utiliser la cam ra dans un environnement poussi reux Retirer la pellicule protectrice recouvrant le verre de la t te d observation Mesures de s cu...

Страница 26: ...surplus de solution doit tre limin en tordant le chiffon avant d essuyer la cam ra Lors de l utilisation de tampons le surplus de solution doit galement tre limin en tordant le tampon avant d essuyer...

Страница 27: ...briqu pour Repr sentant autoris en Europe R paration et mise au rebut La dur e de vie de la cam ra est de cinq ans Aucune pi ce de l appareil ne peut tre r par e ou entretenue Consulter la garantie de...

Страница 28: ...Directives et d claration du fabricant missions lectromagn tiques pour tous les QUIPEMENTS et SYST MES Directives et d claration du fabricant missions lectromagn tiques Cette cam ra est pr vue pour u...

Страница 29: ...r standard Surtension IEC 61000 4 5 1 kV phase phase 2 kV phase terre 0 5 et 1 kV mode diff rentiel 0 5 1 et 2 kV mode commun La qualit de l alimentation de secteur doit tre celle d un environnement c...

Страница 30: ...nsit des champs d metteurs RF fixes telle qu elle est d termin e par l tude lectromagn tique d un site a doit tre inf rieure au niveau de conformit pour chaque plage de fr quences b Des interf rences...

Страница 31: ...metteur W Distance de s paration en fonction de la fr quence de l metteur m De 150 kHz 80 MHz 80 MHz 800 MHz De 800 MHz 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100...

Страница 32: ...ne contient pas de pi ces qui peuvent tre r par es Centre de service la client le DEXIS tats Unis et Canada T l 1 888 883 3947 T l copieur 1 888 833 3947 Courriel support dexis com l ext rieur de l Am...

Страница 33: ......

Страница 34: ...th Penn Road Hatfield PA 19440 USA 1 888 883 3947 www dexis com Manufactured for Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach Germany 49 0 73 51 56 0 Authorized representative 1 012 8910 REV...

Отзывы: