Dewetron HSI-HV Скачать руководство пользователя страница 9

9

HSI-HV

Isolated high voltage amplifier

Input ranges: 

1400 V to 20 V

Bandwidth: 

2 MHz

Isolation: 

1.8 kV

RMS

 line to line 

1.4 kV

RMS

 line to ground

Input resistance: 

10 MΩ

Protection: 

±4 kV surge / burst

Module specifications

HSI-HV

Input ranges

±20 V

1)

, ±50 V

1)

, ±100 V, ±200 V, ±400 V, ±800 V, ±1400 V

1 year accuracy

2)

Range 

     Signal frequency 

             Accuracy

20 V; 50 V 

  DC 

±0.05 % of reading ±60 mV

100 V to 1400 V  

  DC 

±0.05 % of reading ±0.05 % of range

      0.1Hz to 500 Hz 

±0.05 % of reading ±0.01 % of range

>500 Hz to 5 kHz

±0.1 % of reading ±0.05 % of range 

>5 kHz to 50 kHz

±0.4 % of reading ±0.05 % of range 

>50 kHz to 100 kHz

±(0.016*f) % of reading ±0.1 % of range

>100 kHz to 1 MHz

±(0.010*f) % of reading ±1 % of range

>1 MHz to 2 MHz

±(0.014*f) % of reading ±3 % of range

f = signal frequency in kHz

Gain linearity

0.05 %

Gain drift range

Typically 20 ppm/°C (max. 50 ppm/°C)

Off set drift

   20 V to 100 V

   200 V to 1400 V

typical 1.5 mV/°C of range 

max. 4 mV/°C

typical 5 ppm/°C       

max. 20 ppm of range/°C

Long term stability

100 ppm/sqrt (1000 hrs)

Input resistance

10 MΩ || 2.2 pF

-3 dB Bandwidth

2 MHz

Signal delay @ full bandwidth

approx. 390 ns

Filter selection

Push button or software

Filter (lowpass)

100, 300, 1k, 3k, 10k, 30k, 100k, 300 kHz, 1 MHz, 2 MHz

3)

Filter type

Bessel or Butterworth 40 dB/decade

Filter characteristics

  100 Hz to 1 MHz

  2 MHz

Butterworth or Bessel 40 dB/dec (2

nd

 order; ±1.5 dB @ f

0

)

Butterworth 60 dB/dec (3

rd

 order; 0 to -3 dB @ 2 MHz)

Typical SFDR and SNR

   50 V

   400 V

   1400 V

10kHz bandwidth 

100kHz bandwidth 

1MHz  bandwidth 

2 MHz  bandwidth

SFDR          SNR 

SFDR           SNR 

SFDR         SNR 

SFDR          SNR

110 dB          91 dB 

110 dB           82 dB 

94 dB         76 dB 

84 dB          73 dB

110 dB          95 dB  

110 dB           92 dB  

94 dB         82 dB 

84 dB          77 dB

110 dB          95 dB 

110 dB           95 dB 

94 dB         82 dB 

84 dB          77 dB

Typical CMRR

>80 dB @ 50 Hz

60 dB @ 1 kHz

70 dB @ 400 Hz

48 dB @ 10 kHz

Isolation voltage

Line to Ground 1.4 kVrms

Line to Line 1.8 kVrms

Protection

Surge (1.2/50)

Burst (5 kHz)

CAT III 600

CAT IV 300

±4000 V

±4000 V

Output voltage

±5 V

Output resistance

10 Ohm

Output current maximum

35 mA 

CAUTION: do not exceed maximum output current!

Power supply

±9 V

DC

 ± 1%

Power consumption

1.2 W

Power On default settings

Software programable

Special functions

Integrated temperature sensor

Programming interface

RS-485

1)

20 V and 50 V are auxiliary ranges and have a limited bandwidth.

20 V range typically 0.9 Mhz

50 V range typically 1.9 Mhz

2)

Conditions for accuracy: Module temperature is calibration temperature ±5 °C; humidity is 30 to 90 RH;

AC accuracy: the highest fi lter (2 Mhz) has to be activated. f = signal frequency in kHz.

for the 2 year accuarcy multiply all % of range and % of reading values by 1.5.

3)

2 MHz fi lter: exclusively Butterworth 60 dB/decade. Please consider possible bandwidth limitation of further components in the measuring chain, e.g. A/D cards or

signal conditioning mainframe

450134 • DAQP/H

SI

/PAD Modules • Technical Reference Manual • Prin

t

ing version 2.3.4 • November 18, 2020

Содержание HSI-HV

Страница 1: ...supply you with accurate complete and reproducible measurement results for your decision making Look forward to the easy handling and the flexible and modular use of your DEWETRON product and draw upo...

Страница 2: ...protected by copyright All rights are reserved Reproduction adaptation or translation without prior written permission is prohibited except as allowed under the copyright laws All trademarks and regis...

Страница 3: ...se of the information in this document Warranty Information A copy of the specific warranty terms applicable to your DEWETRON product and replacement parts can be obtained from your local sales and se...

Страница 4: ...wing general safety precautions must be observed during all phases of operation service and repair of this product Failure to comply with these precautions or with specific warnings elsewhere in this...

Страница 5: ...ained Keep away from live circuits Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers or shields are for use by service trained personnel only U...

Страница 6: ...com The telephone hotline is available Monday to Friday between 08 00 and 12 00 CET GMT 1 00 and Monday to Thursday between 13 00 and 17 00 CET For the Americas please contact DEWETRON Inc 2850 South...

Страница 7: ...AQ series modules available The CAL KIT contains the required cables software and instructions General module specifications Module dimensions 20 x 65 x 105 mm 0 79 x 2 56 x 4 13 in W x H x D without...

Страница 8: ...e FILTER button again Approx 3 seconds after the last key activity the information will be stored the LED stops flashing and shows the input range again CAUTION Power loss during this time leaves the...

Страница 9: ...erworth 60 dB dec 3rd order 0 to 3 dB 2 MHz Typical SFDR and SNR 50 V 400 V 1400 V 10kHz bandwidth 100kHz bandwidth 1MHz bandwidth 2 MHz bandwidth SFDR SNR SFDR SNR SFDR SNR SFDR SNR 110 dB 91 dB 110...

Страница 10: ...V 102 Push button operation Select range Push the RANGE button several times shortly until the LED displays the desired input range Select filter Push the FILTER button once the LEDs will flash for ap...

Страница 11: ...characteristic could be chosen between Butterworth 2nd order or Bessel 2nd order The highest filter is a 3rd order filter with a guaranteed 3 dB bandwidth of 2 MHz This filter structure is the same fo...

Страница 12: ...der systems may have a fix installed 350 kHz filter The HSI series modules will also work in these systems but the bandwidth will be limited to the system bandwidth IN IN Red Black Signal connection H...

Страница 13: ...ber States relating to electrical equipment designed for use within certain voltage limits amended by the directive 93 68 EEC 89 336 EEC Directive on the approximation of the laws of the Member States...

Страница 14: ...C2 Notes...

Отзывы: