28
DW2-UPS-250-DC
Changing batteries
The batteries of the DW2-250-UPS-DC are exchangeable during operation. To access the batteries, loosen the screws
at the front panel and remove the battery cover of the DW2-UPS-250-DC as shown in the pictures below.
NOTE:
Disconnect supplied devices from the DW2-UPS-250-DC output when they are powered
off! Store the batteries separately if the system is unused for more than two weeks, other
wise they will deep discharge and could get damaged.
Power limitations
The DW2-UPS-250-DC is capable of 250 W output power with 3 batteries installed. By removing a battery the output
power drops to 180 W.
External battery charger (optional)
External battery charger
CH5000A/E/U
Power supply:
Input voltage:
90 to 260 VAC, 24V
Input current:
2.5 A
Mains Cord:
CH5000E - 220 V European 2-pin connector with ground recess
Dimensions (WxDxH):
180 mm x 92 mm x 58 mm
Weight:
ca. 250 g
Mating connector:
5-blade standard battery connector
From time to time, due to the aging process of the batteries, it is necessary to recalibrate the battery in order to retain
the accuracy and reliability of the charge level indicator. This can be archieved with an external battery charger (BAT-
CHARGER) which is optionally available. Another advantage of the BAT-CHARGER is that additional batteries can be
recharged without being in the measurement unit. This allows the measurement unit to run non-stop without being
connected to the power net, thanks to the hot-swap capability of the battery packs.
!
3 batteries installed =
250 W
2 batteries installed =
180 W