Cooling considerations
The intake vent of the DEWE2-PA7 is on the bottom of the chassis, where the cooling exhaust vent for the DEWE2-PA7
is on the rear of the chassis.
CAUTION:
Adequate clearance between the chassis and surrounding equipment or blockages must
be maintained to ensure proper cooling of the chassis power supply as well as the modules
plugged into the chassis!
NOTE:
When installing the DEWE2-PA7 into a 19" cabinet please keep in mind that the unit is
siphoning air from the bottom (see schematic above). Due to this circumstances
5 u are required for one system (4 u = 1 u = cooling)! Otherwise the system
may overheat and could get damaged!
System recovery
For more information regarding a total recovery please refer to the corresponding total recovery technical reference
manual shipped with your DEWE2 system.
cool air in
hot air out
rear
fr
on
t
DEWE2-PA7 side view
4 u (System)
1 u (Clearance)
min. 1 u
(Clearance)
!
!