Maintenance and service
The information in this section is designed for use by qualified service personal.
Service interval
Clean dust from the chassis exterior/interior and exchange filter foam based on the operating environment.
Cleaning
Clean surface of the chassis with dry lint-free cloth.
Use a dry velocity stream of air to clean the chassis interior.
Do not use harsh chemical cleaning agents.
NOTICE
Many components within the chassis are sensitive to static discharge damage. Always wear a ground
wrist strap and service the unit only in static-free environment.
WARNING
Risk of injury
Disconnect all cables before servicing the unit.
Replacing the fuse
The DEWE-CAM-GIGE-SPLIT-01 box is protected by a user replaceable 20 A ATO fuse.
INFORMATION
The fuse has to be replaced by the same type.
Updates
Software updates
NOTICE
The system BIOS is protected by password. Any change in the BIOS may cause a system crash. When
the system is booting, do not press ESC-button on keyboard. This may clear the BIOS settings and cause
system faults.
Any change in the file structure as deleting or adding files or directories might cause a system crash.
Before installing software updates contact DEWETRON or your local distributor. Use only software
packages which are released by DEWETRON. Further information is also available in the Internet (http://
www.dewetron.com).
After power off the system wait at least 10 seconds before switching the system on again. Otherwise the
system may not boot correct. This prolongs also the life of all system components.
!
maintenance and Service
19
DEWE-CAM-GIGE-SPLIT-01 • Technical reference • Printing version 1.1.4 • December 6, 2021
Содержание DEWE-CAM-GIGE-SPLIT-01
Страница 4: ...Preface 4 Notes...
Страница 6: ...TABLE OF CONTENTS 6 Notes...
Страница 22: ...Notes Maintenance and service 22...
Страница 24: ...Certificate of conformity 24 Notes...