![Dewetron DAQP-MULTI Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/dewetron/daqp-multi/daqp-multi_technical-reference-manual_2489960012.webp)
12
DAQP-MULTI / DAQP-THERM
Thermocouple accuracy including CJC error
Type
Standard
Input range
Accuracy
min
[°F] °C
max
[°F] °C
-270 to -200 °C
-454 to -328 °F
[°F] °C
-200 to -100 °C
-328 to -148 °F
[°F] °C
-100 to 0 °C
-148 to 32 °F
[°F] °C
0 to 100 °C
32 to 212 °F
[°F] °C
100 to 400 °C
212 to 752 °F
[°F] °C
400 to 1000 °C
752 to 1832 °F
[°F] °C
>1000 °C
> 1832 °F
[°F] °C
K
DIN EN 60584-1 [-454] -270 [2501] 1372
[17.41] 9.67
[1.82] 1.01
[1.92] 0.51
[0.70] 0.39
[0.79] 0.44
[1.08] 0.6
[1.39] 0.77
J
DIN EN 60584-1 [-346] -210 [2192] 1200
[1.76] 0.98
[1.57] 0.87
[0.85] 0.47
[0.67] 0.37
[0.76] 0.42
[0.92] 0.51
[1.01] 0.56
T
DIN EN 60584-1 [-454] -270
[752] 400
[11.38] 6.32
[1.78] 0.99
[0.99] 0.55
[0.70] 0.39
[0.61] 0.34
-
-
R
DIN EN 60584-1
[-58] -50
[3200] 1760
-
-
[2.30] 1.28
[1.60] 0.89
[1.17] 0.65
[0.95] 0.53
[1.28] 0.71
S
DIN EN 60584-1
[-58] -50
[3200] 1760
-
-
[2.07] 1.15
[1.57] 0.87
[1.21] 0.67
[1.04] 0.58
[1.39] 0.77
N
DIN EN 60584-1 [-454] -270 [2372] 1300
[23.81] 13.23
[2.02] 1.12
[0.97] 0.54
[0.67] 0.42
[0.70] 0.39
[0.86] 0.48
[1.03] 0.57
E
DIN EN 60584-1 [-454] -270 [1832] 1000
[11.00] 6.11
[1.06] 0.87
[0.88] 0.49
[0.65] 0.36
[0.61] 0.34
[0.86] 0.48
-
L
DIN 43710
[32] 0
[1652] 900
-
-
-
[0.65] 0.36
[0.74] 0.41
[0.77] 0.43
-
C
ASTM E988-96
[32] 0
[4190] 2310
-
-
-
[0.88] 0.49
[0.86] 0.48
[1.06] 0.59
[1.69] 0.94
U
DIN 43710
[-328] -200 [1112] 600
[1.67] 0.93
[1.67] 0.93
[0.99] 0.55
[0.70] 0.39
[0.63] 0.35
[0.56] 0.31
-
B
DIN EN 60584-1
[32] 0
[3308] 1820
-
-
-
[54.56] 30.31
[5.47] 3.04
[1.40] 0.78
[0.92] 0.51
= calculated specifi cations, not veryfi ed.
Input ranges and detailed specifications
1)
RTD
Type
Standard
Input range
Current
Accuracy
min
[°F] °C
max
[°F] °C
mA
-200 to -100 °C
-328 to -148 °F
[°F] °C
-100 to 0 °C
-148 to 32 °F
[°F] °C
0 °C to fullscale
32 °F to fullscale
(% of r [°F] °C)
Pt100 (385)
DIN EN 60751
[-328] -200 [1562] 850
0.2
[0.25] 0.14
[0.37] 0.21
0.07
[0.37] 0.21
Pt200 (385)
DIN EN 60751
[-328] -200 [1562] 850
0.1
[0.32] 0.18
[0.48] 0.27
0.10
[0.48] 0.27
Pt500 (385)
DIN EN 60751
[-328] -200 [1562] 850
0.2
[0.61] 0.34
[0.75] 0.42
0.09
[0.75] 0.42
Pt1000 (385)
DIN EN 60751
[-328] -200 [1562] 850
0.2
[0.39] 0.22
[0.52] 0.29
0.09
[0.52] 0.29
Pt2000 (385)
DIN EN 60751
[-328] -200 [1562] 850
0.2
[0.45] 0.25
[0.63] 0.35
0.12
[0.64] 0.36
Pt100 (3926)
[-328] -200 [1562] 850
0.2
[0.25] 0.14
[0.37] 0.21
0.07
[0.37] 0.21
Resistance
Range
Current
Accuracy
[Ω]
[mA]
[% of reading]
[% of range]
1000000
0.005
0.04
1.02
300000
0.015
0.04
0.35
100000
0.05
0.04
0.11
30000
0.1
0.04
0.07
10000
0.1
0.04
0.08
3000
0.2
0.04
0.07
1000
0.5
0.04
0.25
300
1
0.04
0.18
100
1
0.04
0.12
30
2
0.04
0.08
10
4
0.04
0.06
3
5
0.04
0.10
1
5
0.04
0.23
Excitation
[% of reading]
[µA]
0 to 200 µA
0,02
0,05
>0.2 to 5 mA
0,02
1
1)
All accuracy specifications mentioned on this page are 1-year specifications.