DS-CAL
SOFTWARE USER MANUAL
Important
For the Krypton UUT please consult the Krypton Technical Reference Manual for information
about ways of connecting power and data connection and usage of proper cables! Cables for
connecting the Krypton UUT are not included in the DS-CAL-BOX package, see chapter
DS-CAL-BOX.
Illustration 34: Example of connecting the Krypton 8xTH channels using T
‑
type thermocouple cable
9.6.1. Krypton 8xTH, 16xTH
Important
For the calibration of Krypton 8xTH / 16xTH UUT an accessory package KRYPTON TC CAL SET is
required (see chapter KRYPTON-TC-CAL-SET)!
Consider also a usage of supported multifunction calibrator (see chapter Multifunction
calibrator support)!
Krypton TH channels connect directly to the multifunction calibrator using a 1-to-8 or 1-to-16
male-to-male splitter T-type thermocouple cable. Thermocouple cables need to be of hi-precision type.
All channels must be connected and calibrated at once.
Hint
Because of the nature of thermocouple cables having a temperature dependent voltage drop,
shorting of thermocouple input channels is required at one point during calibration. Use the
provided white thermocouple terminators. You will be informed by the DSCalibrator software
during the calibration procedure when shorting is required.
Example of connecting the Krypton 8xTH channels shows Illustration 34.
DS-CAL V20-1
46/67
Содержание DS-CAL
Страница 1: ...DS CAL SOFTWARE USER MANUAL DS CAL V20 1 1...
Страница 20: ...DS CAL SOFTWARE USER MANUAL Illustration 11 Users tab of the DSCalibrator DS CAL V20 1 20 67...
Страница 26: ...DS CAL SOFTWARE USER MANUAL Illustration 17 Prepare adjustment procedure window DS CAL V20 1 26 67...
Страница 55: ...DS CAL SOFTWARE USER MANUAL Illustration 46 Example of connecting the DSI 5A DS CAL V20 1 55 67...