TECHNICAL REFERENCE MANUAL
DS-16xLVDTr
6.3. Typical Half Bridge sensor connection
External Half Bridge completion
Optional internal Half Bridge completion
Important
Input - (pin7) shall be connected to half bridge completion resistor divider assembled with
discrete resistors with the following recommended specifications:
●
Resistance 1kΩ,
●
Tolerance 0.1%,
●
Temperature coefficient 15ppm,
●
Power 0,125W.
This connection is preferred over the connection IN- to GND (so called “noisy GND”).
V1.0
17 / 27