Dewert User Manual Model 2410XL - Updated 05.2021/1 MDR                                                                Page 7 of 12 

4.2 Technical data for hydraulic system 

Manufacturer   

Power-Packer Europa B.V., NL-7575 AT Oldenzaal 

Type 

 

 

Compact MK5 long 

Mode of operation 

Hydraulic cylinder with pump 

 
The hydraulic unit is maintenance free. 
 
 

5. Cleaning instructions 

 
Upholstery covering 

There are two different collections of upholstery covering to choose from: 
 

 

Skai Pandoria Plus (manufactured by Hornschuch/Continental in Germany) 

 

Skai Toronto EN (manufactured by Hornschuch/Continental in Germany) 

 
For cleaning and disinfection, a selection of agents from various manufacturers has been tested for 
compatibility. Please refer to the separate enclosed sheet. 
 

Table frame cleaning 

The plastic-coated table frame and the chrome-plated bars and levers can be cleaned with mild household 
cleaners, if necessary. 
 
Do not use aggressive, abrasive or corrosive agents. Heavily soiled chrome-plated parts can be cleaned with 
a chrome polish (e.g. Sidol). After cleaning, dry the frame with a soft dry cloth. Seal deep scratches and worn 
areas with suitable repair agents to prevent moisture penetration. 
 

Important: 

With the hydraulic pump as well as with gas springs, wipe the piston rod with a soft cloth at regular intervals. 
This prevents dust from entering through the dust lip and preserves the service life of the unit. 
 

 

   Caution: When cleaning, secure the table against unintentional lowering of the lying 

surface. To do so, set all the adjustable sections to the horizontal position and block the foot levers 
against unintentional operation. 

 
Do not clean the table with water jets, a high-pressure cleaner or with a so-called bed washer. Only use moist 
cloths. 
 
 

6. Maintenance and technical inspection 

The device has been designed and manufactured in such a way that it will operate safely over a very long 
period of time if used as intended and in a correct manner. Depending on the conditions of use, place of use 
and care, the expected operating service life is up to 10 years or 100,000 drive cycles (double stroke = 1x up 
and down). 
 

Regular maintenance procedures

 are required to ensure patient, user and product safety. They must be 

carried out every two years at the latest.  
The maintenance procedures can be carried out by qualified personnel/instructed personnel. 
 
The scope of maintenance includes, for example:  
 

 

Thorough visual inspection of all the components, especially the hydraulic pump and connection 

 

Check all swivel joints for completeness. 

 

Check proper fit of the screw connections.  

 

Lubricate the swivel joints and operating levers slightly with low viscosity spray oil, if necessary. 

 

Lubricate the roller guides with a little bearing grease or similar using a brush, if necessary. 
 

Содержание 2410XL/H

Страница 1: ...Safety regulations 2 1 4 Intended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number...

Страница 2: ...g manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons The user must ensure that the device is not accessible to unauthorised persons or cann...

Страница 3: ...strictly prohibited and possibly dangerous The manufacturer cannot be held responsible for damage caused by improper use Patients may only be positioned in preparation for treatment examination by pr...

Страница 4: ...al safety The number of possible pinching points has thus been minimised while remaining ones have been covered or protected with spacers to prevent injury thus ensuring safe and yet simple operation...

Страница 5: ...with infusions 2 4 Movability The table is equipped with easy running comfort castors measuring 150 mm in diameter in a conductive design Central movability By operating a lever on the outside of the...

Страница 6: ...ts of a holding bar and retaining pins that are attached below the bumpers on the head part In addition to the stainless steel bar the paper roll holding bar consists of a spring loaded stainless stee...

Страница 7: ...vents dust from entering through the dust lip and preserves the service life of the unit Caution When cleaning secure the table against unintentional lowering of the lying surface To do so set all the...

Страница 8: ...liable for damages resulting from uncoordinated modifications Only original Dewert spare parts may be used Gas springs are wearing parts whose function may be impaired over the years depending on the...

Страница 9: ...sposal costs meaning you can dispose of the packaging free of charge Product Safety notices o Pay attention to sharp edges and pointed objects o When transporting the table only take hold of the under...

Страница 10: ...similar medical devices have knowledge of its proper use taking into account possible hazards Ideal positioning of patients for the purpose of transport to pre treatment or post treatment locations a...

Страница 11: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Страница 12: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...