Dewert User Manual Model 2410XL - Updated 05.2021/1 MDR Page 7 of 12
4.2 Technical data for hydraulic system
Manufacturer
Power-Packer Europa B.V., NL-7575 AT Oldenzaal
Type
Compact MK5 long
Mode of operation
Hydraulic cylinder with pump
The hydraulic unit is maintenance free.
5. Cleaning instructions
Upholstery covering
There are two different collections of upholstery covering to choose from:
•
Skai Pandoria Plus (manufactured by Hornschuch/Continental in Germany)
•
Skai Toronto EN (manufactured by Hornschuch/Continental in Germany)
For cleaning and disinfection, a selection of agents from various manufacturers has been tested for
compatibility. Please refer to the separate enclosed sheet.
Table frame cleaning
The plastic-coated table frame and the chrome-plated bars and levers can be cleaned with mild household
cleaners, if necessary.
Do not use aggressive, abrasive or corrosive agents. Heavily soiled chrome-plated parts can be cleaned with
a chrome polish (e.g. Sidol). After cleaning, dry the frame with a soft dry cloth. Seal deep scratches and worn
areas with suitable repair agents to prevent moisture penetration.
Important:
With the hydraulic pump as well as with gas springs, wipe the piston rod with a soft cloth at regular intervals.
This prevents dust from entering through the dust lip and preserves the service life of the unit.
Caution: When cleaning, secure the table against unintentional lowering of the lying
surface. To do so, set all the adjustable sections to the horizontal position and block the foot levers
against unintentional operation.
Do not clean the table with water jets, a high-pressure cleaner or with a so-called bed washer. Only use moist
cloths.
6. Maintenance and technical inspection
The device has been designed and manufactured in such a way that it will operate safely over a very long
period of time if used as intended and in a correct manner. Depending on the conditions of use, place of use
and care, the expected operating service life is up to 10 years or 100,000 drive cycles (double stroke = 1x up
and down).
Regular maintenance procedures
are required to ensure patient, user and product safety. They must be
carried out every two years at the latest.
The maintenance procedures can be carried out by qualified personnel/instructed personnel.
The scope of maintenance includes, for example:
•
Thorough visual inspection of all the components, especially the hydraulic pump and connection
•
Check all swivel joints for completeness.
•
Check proper fit of the screw connections.
•
Lubricate the swivel joints and operating levers slightly with low viscosity spray oil, if necessary.
•
Lubricate the roller guides with a little bearing grease or similar using a brush, if necessary.