background image

IPROXX 2 

Troubleshooting 

74828(c)  

27 

5.  Troubleshooting 

 

LED signal/error 

Meaning 

Measure / Action 

No movements can be carried 

out on the end product. 

- The handset is locked. 
 

- The locking mechanism is 

malfunctioning. 
- The power supply or con-

necting cables are defective. 
 
- The battery is not charged or 

not connected. 
- The controller or drive is de-

fective. 

- Use the turnkey to en-

able the handset. 
- Check the locking 

mechanism. 
- Check the power supply. 
- Check the connecting 

cables. 
- Connect and/or re-

charge the battery. 
Please contact your sup-

plier or sales agent. 

The function/CARE LED on the 

handset does not illuminate 

when a key is pressed. 

There is an error in the control 

system. 

Please contact your sup-

plier or sales agent. 

The function/CARE LED on the 

handset illuminates continu-

ously, even when a key is not 

pressed. (Only for single-fault 

protected drive systems) 

There is an error in the control 

system. 

Please contact your sup-

plier or sales agent. 

The battery/power-on LED does 

not illuminate (handset is 

locked; with the exception of 

battery-operated mode). 

There is a handset error. 

Please contact your sup-

plier or sales agent. 

The battery/power-on LED is il-

luminated brightly (handset is 

locked). 

There is a handset error. 

Please contact your sup-

plier or sales agent. 

The status LED is illuminated 

when locked. 

There is a handset error. 

Please contact your sup-

plier or sales agent. 

The status LED does not illumi-

nate when enabled. 

- There is a handset error. 
- There is a controller error. 

Please contact your sup-

plier or sales agent. 

 

 

 

Содержание IPROXX 2

Страница 1: ...System Instructions IPROXX 2...

Страница 2: ......

Страница 3: ...rtOkin is not responsible for damage resulting from failure to observe these instructions changes made to this product which have not been approved by DewertOkin or the use of replacement parts which...

Страница 4: ...wer 12 4 3 Connecting the handset to the drive system 12 4 4 Key functions for the IPROXX 2 handset an example 13 4 5 Key functions for the IPROXX 2 SMP Bluetooth handset 14 4 6 Enabling and disabling...

Страница 5: ...e cannot guarantee that the data images and drawings are complete and correct nor do we accept any liability for the in formation contained therein unless required by law We reserve the right to make...

Страница 6: ...ufacturer against the manufacturer Improper usage Be sure to follow the notices below concerning improper usage You should include them in your product manual in order to inform the users of your end...

Страница 7: ...this illustration Rating plates examples IPROXX 2 IPROXX 2 SMP Bluetooth Model name xxxxx Article number U in 24 29V Input voltage Prod date Calendar week year Serial No Serial number IPX6 Protection...

Страница 8: ...IPROXX 2 SM with a maximum of 10 movement keys 2 stage lock function and one optional lamp key IPROXX 2 SM with a maximum of 10 movement keys 3 stage lock function and one optional lamp key IPROXX 2 S...

Страница 9: ...SMP Bluetooth handset IPROXX 2 SMP Bluetooth handset A Function and movement keys B Function CARE LED C Mounting clip optional D Status LED E Power ON LED F Bluetooth key G Connection cable for the ha...

Страница 10: ...Socket key B Mounting clip optional C Turnkey lock with 2 stage locking function for the IPROXX 2 SM model D Turnkey lock with 3 stage locking function for the IPROXX 2 SM model E Turnkey lock with 3...

Страница 11: ...width x height with mounting clip 170 mm x 60 mm x 51 mm Weight with cable Approx 300 g Ambient conditions for operation storage and transport Transport storage temperature From 20 C to 50 C From 4 F...

Страница 12: ...how to operate this locking mechanism 4 2 Information on battery power If you are operating your system on battery power then observe the following notice NOTICE After it has been used to move your ap...

Страница 13: ...mple IPROXX 2 handset with 11 keys A Function and movement keys B Function CARE LED C Status LEDs only with the IPROXX 2 SMP model D Optional Status LED only for the IPROXX 2 SM SM and SMP models E La...

Страница 14: ...Function and movement keys B Function CARE LED C Status LEDs D Power ON LED E Bluetooth key F Connection cable for the handset G Status LED for Bluetooth connection The IPROXX 2 SMP Bluetooth handset...

Страница 15: ...justments to leg section Upward Downward Reset of back and leg sections Upward Downward Adjustment to stroke Upward Downward Anti Trendelenburg Head upward foot downward Trendelenburg Head downward fo...

Страница 16: ...r of the IPROXX 2 The positions of the turnkey lock are marked by icons Icons on the rear of the handset Icon symbol Meaning Function keys and travel keys are locked for models SM SM and SMP Function...

Страница 17: ...M has a 2 stage locking function The models IPROXX 2 SM and IPROXX 2 SMP have a 3 stage locking function SMP programming mode for nursing staff Functionality of the handset model IPROXX 2 SM Rear of h...

Страница 18: ...n IPROXX 2 18 74828 c Functionality of the handset model IPROXX 2 SM Rear of handset IPROXX 2 SM A Socket key B Handset keys are unlocked C Functions enabled for the patient D Handset keys are locked...

Страница 19: ...ons of the handset model IPROXX 2 SMP Bluetooth Rear of handset IPROXX 2 SMP Bluetooth A Socket key B Handset keys are unlocked C Functions enabled for the patient D Handset keys are locked E Programm...

Страница 20: ...set should be replaced 4 7 2 Unlocking the keys for the IPROXX 2 SM SM and SMP Bluetooth models Rotate the turnkey on the back side of the handset in order to unlock the key functions 1 Insert the soc...

Страница 21: ...keys are now possible 4 8 Display functions The IPROXX 2 handset features a function CARE LED and a battery power on LED The handsets with the lock function also feature a lock status LED They have t...

Страница 22: ...abled a Initial commissioning before smart phone discovery 1 Insert the key into the turnkey lock on the IPROXX 2 SMP Bluetooth 2 Turn the key until the arrow on the turnkey lock is set to Changing mo...

Страница 23: ...he smart phone 8 You can use the app when the turnkey lock is set to either unlocked or to patient mode Function enabled for patients are then shown in the app 9 The status LEDs are illuminated in gre...

Страница 24: ...is confirms the range level setting 4 Turn the key until the arrow on the turnkey lock points to or Then remove the key If the handset is outside of the set Bluetooth range the app s movement keys are...

Страница 25: ...ous setting can be made after clicking on the icon Changing the name of the IPROXX 2 handset 1 Touch the Name button The Rename device window opens 2 Enter a new name for the device and save it Changi...

Страница 26: ...circle diagram 4 9 7 Resetting to factory default settings Any paired device and smart phone profiles will be deleted when you reset to the factory default set tings 1 Insert the key into the turnkey...

Страница 27: ...l system Please contact your sup plier or sales agent The function CARE LED on the handset illuminates continu ously even when a key is not pressed Only for single fault protected drive systems There...

Страница 28: ...The corresponding locking LEDs for the affected key pairs flash yellow The handset must be switched off before it can be switched on and operated again Pull the connec tor wait 30 seconds and then re...

Страница 29: ...for any signs of damage Check the connecting cables for pinching or shearing Also check the strain relief and kink protec tions mechanisms in particular after any mechanical load At least every six m...

Страница 30: ...WFD as of 12 12 2008 7 2 Handset components The IPROXX 2 handset consists of electronic components cables and metal and plastic parts You should observe all corresponding national and regional enviro...

Страница 31: ...tives Electromagnetic Compatibility Directive 2004 108 EC Low Voltage Directive 2006 95 EC RoHS Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of...

Страница 32: ...ent RoHS Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment Applied...

Страница 33: ...t 100 A EN 60601 1 Section 17 EMC The following standards and norms were used in accordance with EN 60601 2 52 IEC 60601 2 52 third Edition Particular requirements for the safety and essential perform...

Страница 34: ...lation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the inter...

Страница 35: ......

Страница 36: ...DewertOkin GmbH Weststrasse 1 Kirchlengern 32278 Germany Tel 49 0 5223 979 0 Fax 49 0 5223 75182 http www dewertokin de Info dewertokin de ID No 74828...

Отзывы: