2
1 Device Description
Legend
(Description of individual parts: see
enclosed leaflet)
1
Headgear
2
Forehead cushion
3
Fine adjustment component
4
Coarse adjustment component
5
Forehead support
6
Ports for pressure measurement
7
Oxygen port cap
8
Rotating sleeve
9
Elbow
10
Retaining ring
11
Mask frame
12
Mask cushion
13
Headgear clip
Intended Use
The nasal mask is intended for prescription use
to be used during nasal CPAP or Bi-level
Positive Pressure therapy for adult patients
(> 66 pounds or 30 kg). The nasal mask is
offered in two versions:
EasyFit Nasal Mask, Silicone
EasyFit Nasal Mask, Gel
Before use by the patient
Patients must be instructed in the use of the
mask by qualified personnel before therapy
begins.
2 Safety instructions
Warnings and cautions
For your safety, please observe the following
points:
Warning
• Please read these instructions for use very
carefully. They are a part of the product
and should be available at all times.
• This mask should be used only with CPAP
or bi-level systems recommended by your
physician or respiratory therapist.
• If oxygen is used with the CPAP or bi-level
system, the oxygen flow must be turned off
when the CPAP or bi-level machine is not
operating.
Explanation of the warning
: When the
CPAP or bi-level device is not in
operation, and the oxygen flow is left on,
oxygen delivered into the CPAP tubing
may accumulate within the CPAP or bi-
level machine enclosure. Oxygen
accumulated in the CPAP or bi-level
machine enclosure will create a risk of
fire. This warning applies to most types of
CPAP or bi-level systems.
• The mask is designed for therapy pressure
between 4 and 30 cm H
2
O. The nasal
mask must not be used at higher or lower
therapy pressures.
• Patients with limited spontaneous
respiration must be monitored at all times.
The appropriate negative pressure/leak
alarm on the ventilation device must be
activated.
• Patients with obstructive and restrictive lung
diseases should be examined on an individ-
ual basis during use of the mask, for exam-
ple with regard to re-inhalation of CO
2
.
• The mask is equipped with an integral
exhalation valve. Do not connect any
additional exhalation valves as desired
therapy pressure may not be reached.