eN - 10
A-DV56
DeVilbiss Masks
FlexSet
®
Masks
Serenity
®
Masks
Large, Silicone
9354L
Large, Silicone
9352L
Medium, Silicone
9354D
Medium, Silicone
9352D
Small, Silicone
9354S
Small, Silicone
9352S
Large, Gel
9354GL
Large, Gel
9352GL
Medium, Gel
9354G
Medium, Gel
9352G
Small, Gel
9354GS
Small, Gel
9352GS
Large, ComfortTouch
™
Silicone
9354LR
Large, ComfortTouch Silicone
9352LR
Medium, ComfortTouch Silicone
9354DR
Medium, ComfortTouch Silicone
9352DR
Small, ComfortTouch Silicone
9354SR
Small, ComfortTouch Silicone
9352SR
Large, ComfortTouch Gel
9354GLR
Large, ComfortTouch Gel
9352GLR
Medium, ComfortTouch Gel
9354GR
Medium, ComfortTouch Gel
9352GR
Small, ComfortTouch Gel
9354GSR
Small, ComfortTouch Gel
9352GSR
EasyFit
®
Nasal
Available in Silicone (S, M, L, XL), Gel (S, M, L) or SilkGel (S, M, L). Contact
your equipment provider
EasyFit Full Face
Available in Silicone (S, M, L, XL), Gel (S, M, L) or SilkGel (S, M, L). Contact
your equipment provider
MaiNTeNaNce
DaNGer
Electric shock hazard-Do not attempt to open or remove the cabinet; there are no user-serviceable components
inside. If service is required, contact your equipment provider for instructions on obtaining service. Opening or
attempting to service your device will void the warranty.
WarNiNG
Disconnect supply before servicing.
Pressure Accuracy–The IntelliPAP is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer’s
directions.
NOTE–Some countries require periodic servicing and calibration of this kind of medical device. Contact your equipment
provider for further information.
Standard Air-Inlet Filter–Check the filter every 10 days and clean it if necessary. Replace the filter every 6 months or
sooner if it is damaged. The standard filter is designed to filter particles larger than 3.0 microns in size. The IntelliPAP
MUST have the standard filter in place during operation.
Optional Fine Particle Filter–Check the filter every 10 days and replace it if it is dirty or damaged; otherwise replace the
filter every 30 days. The optional fine particle filter is designed to filter particles as small as 0.3 microns in size.
cleaNiNG
WarNiNG
To avoid electric shock, always unplug the power cord from the wall outlet power source before cleaning.
CAUTION–Never rinse or place the device in water. Never allow liquids to get into or around any of the ports, switches or air
filter; doing so will result in product damage. If this occurs, do not use the device. Contact your equipment provider for service.
Device cabinet
1. Unplug the IntelliPAP from the power source. Wipe the cabinet with a clean, damp cloth every few days to keep the
cabinet dust free.
2. Allow the device to dry completely before plugging it into a power source.
Filters
Air-Inlet Filter
1. The air-inlet filter should be checked every 10 days and cleaned as necessary. The filter is located in the rear of the
device.