M A I N T E N A N C E - D V 6 H H
MAINTENANCE AND TESTING
A. CLEANING
Humidifier Water Chamber
NOTE: DeVilbiss recommends the use of distilled water in the DV6
water chamber.
NOTE: Use distilled water to avoid mineral deposit build up inside the
chamber.
NOTE: DeVilbiss does not condone the use of essential oils or other
additives in the distilled water.
Humidifier water chamber should be cleaned daily.
1. Press down on the disassembly lever latch and pull the chamber
from the base (cradle).
2. Press the release bar toward the release latch and lift the lid to
separate chamber.
Optional: Remove sealing gasket from chamber lid only if desired.
Gasket does not need to be removed for cleaning.
3. Hand wash in a solution of warm water and mild dishwashing
detergent. Rinse thoroughly and allow the parts to air dry before
reassembly.
NOTE: The disassembled chamber may also be washed once per
week in a dishwasher on the delicate or glassware cycle, top shelf
only.
NOTE: If the sealing gasket was removed for cleaning, replace
gasket into its groove around the top of the chamber ensuring that it is
snapped under the 12 ribs.
CAUTION: Inspect the sealing gasket during each cleaning to ensure
that it is not damaged and is properly seated under the ribs and in the
groove around the entire chamber.
4. Place the chamber lid’s tabs into the hinge slots on the chamber
base (cradle) and press the front of the lid until it clicks into place.
Return the chamber to the base (cradle).
Humidifier Base (Cradle)
The humidifier base (cradle) should be cleaned weekly.
1. Stop the CPAP blower and remove the unit from its power
source. Allow the system to cool for at least 10 minutes.
2. Disconnect tubing and remove the water chamber from the
humidifier base (cradle) before detaching the base from the
CPAP device.
3. Wipe the base (cradle) with clean damp cloth. Allow it to dry
completely before reattaching to the CPAP device, refitting the
tubing and chamber. Return to its power source for continued
use.
B. MAINTENANCE
Humidifier Base (Cradle) – Clean the base weekly.
Humidifier Water Chamber (single-patient use) – Clean the chamber
daily. The chamber should be replaced as needed.
C. TESTING
Heater Plate Temperature Test
Additional equipment needed: Thermometer
1. Using an empty water chamber, apply AC power to the unit and
start the blower.
2. Enter Clinical mode.
3. Go to CPAP Settings, select the Humidity menu and ensure that
a setting of 1 is enabled.
4. Allow the heat to stabilize for 3 minutes before removing the
water chamber and laying thermometer on the heater plate.
5. After 1 minute verify that the temperature is 33˚ C (+/- 2˚)
6. Return water chamber to heater base (cradle).
7. Go to CPAP Settings, select the Humidity menu and ensure that
a setting of 5 is enabled.
8. Allow the heat to stabilize for 3 minutes before removing the
water chamber and laying a thermometer on the heater plate.
9. After 1 minute verify that the temperature is 65˚ C (+/- 2˚)
10. Correct all settings to the original preferred or prescribed settings.
WARNING
To avoid burns, never touch the heater plate.
NOTE: If there is an issue with the temperature of the heater plate,
proceed to the troubleshooting section.
Pulse Test - for optional PulseDose® Valve Module on Heated
Humidifier
Additional equipment required: USB to Micro B USB CPAP to PC
cable (DV63D-615), outlet cap (DV51D-620) and a computer running a
terminal program (Microsoft Windows HyperTerminal or equivalent).
1. Connect USB to Micro B cable from the computer to the CPAP
device’s USB port.
2. Connect an outlet cap to the heated humidifier’s air supply port.
3. Apply power to the device.
4. Enter the following in the terminal program:
• Mo=2 - Program returns 2
• Hs=1 - Program returns 01
5. Observe that the Heater ON / Hot Surface Indicator light is now
active.
6. Lift the outlet cap to a slight angle and observe that the pulse
valve is rotating. If the cap is too open, the device will not trigger
the pulse.
NOTE: If there is an issue with operation of the PulseDose valve,
replace PulseDose valve module.