EN - 16
SE-DV57-1
Dynamic Short-Term Pressure Accuracy with and without DV5HH Humidifier (Per
ISO 17510-1:2007)
Waveform
Volume [mL]
Breath Rate [min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
[Pk-Pk cmH
2
O]
Sinusoidal
cycle with I:E =
1:1
500
10, 15, 20
3.0
0.5
9.0
0.5
14.0
1.0
20.0
1.0
25.0
1.0
Maximum Flow-Rate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH
2
O
9.0 cmH
2
O 14.0 cmH
2
O 20.0 cmH
2
O 25.0 cmH
2
O
Measured pressure at the patient
connection port (cmH
2
O)
3.0
8.9
13.0
19.0
24.0
Average flow at the patient connection port
(l/min)
95.0
160.5
172.2
158.5
139.7
Long Term Static Pressure Accuracy ................................................................................................................. +/- 0.5 cmH
2
O
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ......................................................................................IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.