11
•
Push the active Biopsy Button only once to initiate the biopsy cycle. To
interrupt the biopsy cycle, push the Biopsy Button again when the cutter is
retracting from position 1 to 3 (See Illustration 6) or when the cutter is
advancing from position 3 to 2. When the biopsy cycle is interrupted, the
cutter will retract to position 3. The Biopsy Indicator LED flashes orange.
To complete the biopsy cycle, push the Biopsy Button.
Illustration 6:
CAUTION:
Moving parts. Ensure nothing is blocking the needle aperture at this time.
10. Rotate the aperture, via rotation of the Holster and Probe, for the next sample
acquisition and repeat step 6 as necessary to obtain additional samples.
•
To avoid bending the needle, do not apply excessive pressure when
positioning within the breast or axilla.
•
Verify collection of desired specimen using post biopsy imaging of the
patient as necessary.
11. If proceeding WITHOUT using a biopsy site identifier, once the final tissue
sample is obtained, remove the device and care for the incision site as
appropriate. Advance to Step 10.
12. If proceeding WITH using a biopsy site identifier, once the final tissue sample is
obtained, disconnect the Integrated Coaxial Cannula and remove the Probe,
leaving the cannula in place, by pressing the Coaxial Release Button (See
Illustration 7) to facilitate marker placement in the biopsy cavity.
Illustration 7: