CPAP SILICONE 
FULL FACE MASK

INtENdEd USE

The CPAP/VPAP Mask is intended to be used for patients who have 

been prescribed CPAP or VPAP therapy and are under the specific 

direction of a physician. It is used in conjunction with continuous 

positive airway pressure (CPAP/VPAP) devices in home, hospital and 

Institutional environments. 

CONtrAINdICAtIONS

This mask may not be suitable for patients with the following conditions: 

impaired cardiac sphincter function, excessive reflux, impaired cough 

reflex and hiatal hernia. It should not be used if the patient is 

unresponsive, obtunded, unable to remove the mask, or uncooperative.

WArNINgS

•  This mask should only be used with CPAP systems recommended 

by your physician or respiratory therapist.

•  Do not use this mask unless the CPAP or VPAP system is turned 

on and operating properly.

•  Never block the mask’s vent holes or exhalation port. 

Note: CPAP systems are intended to be used with special masks 

with vent holes in either the connectors or the mask that allow 

continuous flow out of the mask. When the CPAP device is turned on 

and functioning properly, new air from the CPAP device flushes the 

exhaled air out through the mask’s exhalation port. However, when the 

CPAP device is not operating, enough fresh air will not be provided 

through the mask and exhaled air may be rebreathed. Rebreathing of 

exhaled air for several minutes can in some circumstances lead to 

suffocation.

•  If oxygen is used with the CPAP system, the oxygen flow must be 

turned off when the CPAP device is not in operation. 

Note: When the CPAP device is not in operation and the oxygen flow 

is left on, oxygen delivered into the ventilator tubing may accumulate 

within the CPAP device’s enclosure and create a risk of fire. 
Note: At a fixed flow rate of supplemental oxygen, the inhaled 

oxygen concentration will vary depending on the pressure setting, 

patient’s breathing pattern and leak rate.

•  Do not smoke if oxygen is in use.

•  Contact a health care professional in the event chest discomfort, 

shortness of breath, stomach distention, belching, or severe 

headache are experienced.

•  Avoid eating or drinking three hours before using this mask to 

minimize the risk of vomiting during sleep.

•  Do not use this mask if you are taking medication that may cause 

vomiting. 

CautioN: At low CPAP pressures the flow through the exhalation 

port may be inadequate to clear all exhaled gas from the tubing and 

some rebreathing may occur. 

•  Must only be used by persons who have received adequate 

training from qualified medical personnel.

•  Used in combination with other medical devices might alter the 

performance of the mask.

CAUtION

USA federal law restricts this device to sale by or on the order of a 

physician. 

The use of bleach, chlorine, alcohol or aromatic-based solutions may 

cause hardening and reduce the life of the product.

FEAtUrES

•  One piece molded silicone rubber; small, light and comfortable.

•  Clear mask allows a good view of mouth and nasal area.

•  Low dead space results in low CO

2

 build up.

•  Dual swivel 360° rotating elbow and tubing connector.

•  Headgear with Velcro fasteners provides maximum versatility and 

movement.

CLEANINg INStrUCtIONS

•  Detach headgear from mask.

•  Handwash gently with mild soap and warm water (30˚C), rinse the 

mask and components thoroughly. Note - Headgear should only 

be hand washed with mild soap by gently rubbing in warm water at 

30°C. No washing machine, no ironing.

•  Allow mask and components to air dry before reusing. Do not 

place in direct sunlight.

•  Replace mask if there are cracks, tears or excessive discoloration.

•  The D100 is dishwasher safe if placed in top rack. 

•  important Note: When removing the mask for cleaning, there 

is no need to unfasten the Velcro tabs. Simply unclip the lower 

buckles and remove the upper headgear straps from the slots 

in the mask frame. Replace after cleaning.

•  To remove the mask and headgear, simply unbuckle both 

buckles (A3 & A4) from the mask frame and pull off the mask.

•  To reuse the mask and headgear, simply place upper headgear 

over head and hook both buckles.

•  This product does not require maintenance.

Please follow assembly instructions below:

DeVilbiss

®

 is a registered trademark of 

DeVilbiss Healthcare. 
© 2013 DeVilbiss Healthcare LLC. 08.13 

All Rights Reserved.  A-D100F Rev. B

EN

 

ReF:

Date of  

Manufacture:
Use By:

Consult instructions for use

EC

REP

Manufactured for  

DeVilbiss Healthcare LLC 

100 DeVilbiss Drive 

Somerset, PA 15501-2125 • USA 

800-338-1988 • 814-443-4881

eu Representative 

mdi europa GmbH 

Langenhagener Str. 71

D-30855 Langenhagen, Germany • Tel: +49 511 39089530 

Email: [email protected] • internet: www.mdi-europa.com

DeVilbiss Healthcare GmbH 

Kamenzer Straße 3 

68309 Mannheim GeRMANy 

+49-621-178-98-230

Manufactured by HSiNeR Co., Ltd 

312, Jhongshan Rd., Shengang Dist 

Taichung City 429, Taiwan 

Tel: 886-4-25152480 • Fax: 886-4-25152482 

Email: [email protected] • Web Site: http://www.hsiner.com

2

Identify headgear fastening positions 

on the mask.

5

Pull the upper headgear strap A2 

through the slot B2 on the mask, 

making sure the Velcro faces outward 

Fasten the Velcro tab.

3

Pull the upper headgear strap A1 

through the slot B1 on the mask, 

making sure the Velcro faces outward. 

4

Fasten the Velcro tab.

Place headgear with Velcro side 

down on a flat surface. Detach 2 

Velcro tabs on A1 & A2.

1

Place the headgear over your head 

and adjust both upper headgear 

straps to a comfortable tension.

6

Hook the plastic buckle A3 onto B3 

and A4 onto B4 on mask frame.

7

Adjust both lower headgear straps A3 

& A4 to a comfortable tension.

9

Readjust both upper headgear straps 

A1 & A2 to a comfortable tension.

8

ensure the overhead strap is in 

proper position to avoid slippage.

10

a1

a2

a3

a4

B1

B2

B3

B4

LATEX

Not made with natural rubber latex.

Therapy Pressure (cmH2O)

Exhaust Flow Rate at Given Therapy Pressure

Exhaust F

lo

w

R

at

e (LP

M)

 

Storage relative humidity: < 85%

 

Operation temp: -18°C - 50°C

  Storage temp: -40°C - 60°C

ReSiStaNCe 

(cmH

2

o)

50LPM 100LPM

D100F

0.24

0.62

MoDeL

VoLuMe (ML)

D100F-S

296.6

D100F-M

339.4

D100F-L

367.8

Expected Service Life

•  Mask Frame Components = 6 months 

 

•  Headgear = 3 months 

 

•  Mask Cushions = 3 months

Size:  Assembled Mask (without Headgear): 161mm (H), 90mm (W),  

69mm (D)

Weight:  Mask incl. Headgear: 122g; Headgear: 31g

Materials used:

   Mask frame:  Polycarbonate (PC))

Forehead Pad:  Silicone

   Headgear:  Nylon + Spandex

Cushions:  Silicone

The mask system does not contain latex materials. This product does not 

contain PVC or Phthalates.