0065-0834-0M Rev C
•
8500/8550 Service Manual
7
E. Final Test
1. Perform Corner Test using 1/4 capacity weights (positions 1-6). Use tolerances listed
in the Calibration Table below.
2. Perform Linearity test. Check both ascending and descending. See table below.
3. Install indicator cover (MV1).
4.
For kg units calibrated using lb weights, reset the Cap = 300 after calibration.
8500, 8550, 8500KGEU, 8550KGEU Linearity Check Table
8500 / 8550
Applied
Weight
Tolerance
WEIGHT
POS
2 5 4
7 8
1 6 3
0
0
*200 lb
199.8 lb To 200.2 lb
1,2,3,4,5,6
400 lb
399.8 lb To 400.2 lb
1+5+3
Then
2+6+4
500 lb
499.8 lb- To 500.2 lb+
7 Then 8
750 lb
749.4 lb To 750.6 lb
7+8
Weight Positions
For Corner Test
8500KGEU / 8550KGEU
Applied
Weight
CALIBRATING WITH lb
WEIGHTS
Tolerance
WEIGHT
POS
0
0
*
200 lb
(90.71 kg)
90.6 kg To
90.8 kg
1,2,3,4,5,6
400 lb
(181.43 kg)
181.3 kg To 181.5 kg
1+5+3
Then
2+6+4
500 lb
(226.79 kg)
226.6 kg To 226.9 kg
7 Then 8
*
Use these weights for Corner Test
750 lb
(340.19 kg)
339.9 kg To 340.47 kg