The solo Series Digital Clinical Scales are equipped with software that tests various portions
of the scale's circuitry and verifies proper operation. Should a problem be detected, a
message will be displayed alerting the operator to that condition. The following lists the
messages displayed and their meaning:
DISPLAY MEANING
99
This will be displayed for a few seconds when the scale cannot establish
zero, and then turn the scale off. Refer to the TROUBLESHOOTING
section.
!6;`
Displayed when the batteries are near the point they need to be replaced.
#$
Weight on scale exceeds 570 lb (260 kg).
TROUBLESHOOTING
Your scale has been designed to provide you with years of trouble-free operation. Despite
this, troubles sometimes happen. Before calling for service assistance you should make
some initial checks to verify that a problem does exist. The following describes several types
of symptoms along with suggested remedies.
PROBLEM POSSIBLE
SOLUTIONS
The display
does not turn on
Battery Operation:
y
Check if the batteries are installed and correctly.
y
Are the batteries discharged? Remove the old batteries and replace
them with new ones.
AC Operation:
y
Is the 9V AC/DC power adapter fully inserted into the wall receptacle?
y
Check the wall receptacle for proper AC power. Try another electrical
appliance in the same receptacle. Does it work?
y
Check the circuit breaker.
y
Has there been a power failure?
Incorrect weight
displayed
Make sure the scale platform is not touching an adjacent object.
Have proper operation procedures been followed?
The display
shows
99
and shuts off
If the platform is not empty (the patient is on the scale) when powering
up, the display will show
99
and turn off. Remove any weight from the
platform (have the patient step off the scale) and power up the scale
again. If the scale continues to show
99
and turns off, consult your
scale service representative.
The display
shows a moving
small
6
If the display shows a moving small
q
when powering up, make sure the
scale is stable and wait for a moment until the display shows
.
.
If the scale continues to show a moving small
q
, please consult your
scale service representative.
Содержание solo
Страница 1: ...solo Series Digital Clinical Scale Owner s Manual Model SOLO Models SOLO RI PD100 ...
Страница 2: ...8525 M336 O1 Rev A y SlimPRO Operation 2 ...
Страница 35: ...0044 0391 0M Rev F y solo Series Owner s Manual 31 PARTS IDENTIFICATION CONT Display and Column Sub Assembly ...
Страница 36: ...0044 0391 0M Rev F y solo Series Owner s Manual 32 PARTS IDENTIFICATION CONT Base Sub Assembly ...
Страница 37: ...0044 0391 0M Rev F y solo Series Owner s Manual 33 PARTS IDENTIFICATION CONT Height Rod Sub Assembly ...