![Detecto APEX-RI Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/detecto/apex-ri/apex-ri_owners-manual_2487754022.webp)
3300-0329-0M Rev C
y
apex Portable Scale Owner’s
16
SCALE SETUP, CONT.
'9#
(OIML)
This setting controls the OIML (European) specific requirements.
The display will show
'9#
and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to toggle the OIML selection between
(*
or
06
and
then press the
key to save it and proceed to the next prompt.
Allowable settings are
(*
for EU or
06
for US.
$'#
(Power Up Zero)
The display will show
$'#
and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to toggle the power up zero between
(*
or
06
and
then press the
key to save it and proceed to the next prompt.
This setting controls zeroing of the scale on power up. If enabled, the scale will be reset to
zero automatically on power up. Allowable settings are
(*
or
06
.
#;9
(Zero Tracking)
The display will show
#;9
and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value, and
then press
the
key to save it and proceed to the next prompt.
This setting controls the automatic zero tracking of the scale. Allowable values are 0 to 10,
indicating the number of half-divisions above or below zero that the scale will attempt to
maintain the zero position.
NOTE:
Select 0 (zero) to disable zero tracking.
*9
(Com1 Mode)
The display will show
*9
and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to toggle the Com1 Mode between
60;
(continuous output),
$#!!
(weight on demand) and
:8#;
(Welch-Allyn Spot LXi
interface), then press the
key to save it and proceed to the next prompt.
This setting controls the output for Com 1. Allowable settings are
60;
,
$#!!
or
:8#;
.
If
60;
(continuous output) was selected for Com 1, the scale will continuously transmit
weight data.
If
$#!!
(Weight-On-Demand) was selected for Com 1, and the scale is connected to a
computer for transmission of weight data to a PC-based EMR (electronic medical record)
software program, it will transmit a single set of weight data each time the computer sends a
weight request “ENQ” (hex 05) or a SMA weight request “W”.
If
:86;
(Welch-Allyn Spot LXi interface) was selected for Com 1, the scale will
automatically interface to a Welch-Allyn Spot LXi interface device to transmit weight data.