Model 610-N4X-SA

Adding/Removing Control Modules 

When adding or removing Control Modules it is imperative that the backplane jumpers are correctly removed
or added to insure proper operation of the fault alarm circuitry (see Figure 7).  When removing a Control
Module, the jumpers for Fault (JPx-G, H, and I) directly to the left of the module being removed, must be 
installed to allow proper signal flow and fault operation.  When a Control Module is being added to the 
configuration, the jumpers (JPx-G, H, and I) directly to the left of the module being installed, must be removed
to insure that the signal does not bypass the new module.  Incorrect placement of jumpers will cause the fault
alarm circuit to operate incorrectly.

CH4

CH3

CH5

CH6

Jumpers installed for Control

Modules not installed in unit.

JP9

A

B

C

D

E

F
G

H

I

JP8

A

B

C

D

E

F
G

H

I

JP7

A

B

C

D

E

F
G

H

I

JP6

A

B

C

D

E

F
G

H

I

Jumpers removed for

Control Modules 7 and 8.

Fault Jumper positions

G, H, and I.

1

2

3

4

5

6

7

9

9

10

11

12

13

14

1

2

3

4

5

6

7

9

9

10

11

12

13

14

1

2

3

4

5

6

7

9

9

10

11

12

13

14

1

2

3

4

5

6

7

9

9

10

11

12

13

14

FAULT Output to Fault Relay

Figure 7 

Backplane Configuration Jumpers

Model 10 Control Modules

A Model 10 control module may be changed by simply loosening its mounting screw and sliding the module
out of its card cage.  See the Model 10 Instruction Manual for more information on the Model 10 controller.

Circuit Breaker/Power Switch

If the circuit breaker should trip due to a high current or over voltage, the circuit breaker can be reset by
cycling the breaker ‘OFF’ and back ‘ON’.  If the breaker trips again, a problem has occurred that will need to
be alleviated before power can be restored.

Replacement of the circuit breaker is accomplished by disconnecting power from the unit, disconnecting the
wires going to the breaker, and removing the breaker from the Din-Rail.  Replace the breaker and reconnect
the wires to the new breaker. 

Power Supply

Replacement of the power supply is accomplished by, removing the 10 screws that secure the 610-N4X-SA
face plate, disconnect the two power supply plugs from their respective headers, remove the three screws 
holding the power supply bracket to the Motherboard.  Remove the four screws that hold the power supply
bracket to the power supply; disconnect the wires from the power supply and reassembly with the new power
supply.  Reference Figure 6 for wiring details.

610-N4X-SA Instruction Manual

Rev. 0.1

Page 7 of 8 

Содержание 610-N4X-SA

Страница 1: ...ouston Texas Gas Detection Control System www detcon com Houston Texas Detcon Inc Operator s Installation and Instruction Manual Covers 3 and 5 relay Stainless and Fiberglass Enclosures DETCON Inc 320...

Страница 2: ...Model 610 N4X SA This page left intentionally blank 610 N4X SA Instruction Manual ii...

Страница 3: ...2 1 Start Up 6 3 0 Maintenance Repair 6 4 0 Spare Parts List 8 5 0 Warranty 8 Table of Figures Figure 1 Mounting 3 Figure 2 Motherboard Schematic Wiring Diagram 4 Figure 3 Component Layout 5 Figure 4...

Страница 4: ...ank Shipping Address 3200 A 1 Research Forest Dr The Woodlands Texas 77381 Mailing Address P O Box 8067 The Woodlands Texas 77387 8067 Phone 888 367 4286 281 367 4100 Fax 281 292 2860 www detcon com s...

Страница 5: ...r Discrete output terminal strips located on the controller motherboard are provided for sensor terminations The 4 20 mA outputs for remote recording devices and an RS 485 serial ModBus output are loc...

Страница 6: ...le applying power to annunciators attached to alarm 1 H2S or LEL Low Alarm 1 conditions can be Acknowledged and Reset via the RESET ACK Switch on the front panel of the unit Alarm 2 H2S LEL High Durin...

Страница 7: ...MODEL 10 MODEL 10 MODEL 10 MODEL 10 MODEL 10 MODEL 10 Model 610 N4X www detcon com d e t c o n I n c Figure 1 Mounting 2 Refer to installation and wiring detail of remote mount sensor assemblies as d...

Страница 8: ...6 5 4 3 2 1 14 13 12 11 10 9 8 7 6 5 4 3 2 1 CH1 P12 P16 CH1 4 20mA FAULT VDC Alarm 2 Alarm 1 Fault Out COM NC ALM1 ALM2 GND Shld B A VDC VAC GND VDC N L1 RED BLK N L1 Sensor NO COM NO NC COM NO NC m...

Страница 9: ...D ALM COIL COMMON ALARMS ALM 1 FAULT VDC ALM 2 RS 485 ALM POWER L1 N N GND A B Shld OUT 4 20 FAULT NO C NC NO C VDC IN ALM RESET L1 SENSOR ALM2 NC NO C NC mA CH1 CH2 CH3 CH5 CH4 CH6 VAC IN P4 P5 P7 P7...

Страница 10: ...sure Figure 3 and L1 N EARTH in Figure 6 If applicable connect 220VAC The power supply will accept both voltages 9 Connect Earth Ground to Terminal TB1 1 and Neutral to terminal TB1 5 Figure 6 7 8 9 5...

Страница 11: ...gure 7 Backplane Configuration Jumpers Model 10 Control Modules A Model 10 control module may be changed by simply loosening its mounting screw and sliding the module out of its card cage See the Mode...

Страница 12: ...ircuit Breaker 500 505000 000 Low Voltage Cut off PCB If applicable 3423 2 12V 2 9Ah Battery If applicable 5 0 Warranty Detcon Inc as manufacturer warrants under intended normal use each new Model 610...

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Отзывы: