DISaSSEMBLy aND CLEaNINg pROCEDURE

The PIRVOLHV should be inspected periodically to ensure 

that its performance is not impaired by fouled optics or by 

clogging of the sample draw cup or hydrophobic filter.  

Inspection and/or periodic maintenance involves three 

different areas.

IMPORTANT

Remove power before disconnecting and 
removing the PIRVOLHV for maintenance.

Sample Draw Cup.  

Perform a visual inspection of the 

sample draw cup and associated plumbing, and clean 

as required.  External filtration of sample stream is 

recommended.

hydrophobic Filter.  

The flow of gas through the 

filter can be inhibited by an accumulation of mud or 

other contaminants.  To inspect the hydrophobic filter, 

disassemble the PIRVOLHV as described below.  If the 

filter appears to be fouled, replace it.

Optics.  

Cleaning of the optical surfaces is required only 

if an optical fault is indicated (1.0 mA current output signal 

from the PIRVOLHV).  This procedure is most easily 

accomplished on a bench.

Required materials:  Clean, flat work surface, foam tipped 

swabs (no cotton), isopropyl alcohol, screwdriver.

CAUTION

The PIRVOLHV contains semiconductor devices 
that are susceptible to damage by electrostatic 
discharge.  An electrostatic charge can build 
up on the skin and discharge when an object is 
touched.  Therefore, use caution when handling 
the device, taking care not to touch electronic 
components or terminals.  If the electronics 
assembly is removed, it should be placed in an 
anti-static bag or box while stored or transported.  
A static safeguarded work area is highly 
recommended (if available) for disassembly and 
cleaning of the PIRVOLHV.

1.  Remove the sample draw cup from the front of the 

PIRVOLHV by pulling it straight off the electronics 

assembly (no tools needed).  See Figure 12.

2.  Perform a visual inspection of the sample draw 

cup and associated plumbing, and clean as 

required.  Also inspect the two O-rings inside the 

sample draw cup and clean, lubricate or replace 

as required.

3.  Loosen the two captive screws on top of the 

mirror assembly (Figure 12) and remove the 

mirror assembly and hydrophobic filter from the 

electronics assembly.

4.  Thoroughly douse the interior of the mirror 

assembly as well as the foam tipped swab 

with isopropyl alcohol.  Use the swab to gently 

cleanse the surfaces of the reflecting mirrors 

inside the mirror assembly.  After cleaning with 

the swab, flush out the mirror assembly using 

a liberal amount of isopropyl alcohol.  Tip the 

mirror assembly with mirror openings downward 

to remove accumulated isopropyl alcohol and 

particle contaminants.  Repeat the alcohol flush 

to remove any remaining contaminants.  Allow the 

mirror assembly to air dry in a dust-free location.

IMPORTANT

Do not insert any sharp object into the mirror 
assembly.  Scratching of the mirrors will void the 
PIRVOLHV warranty.  Do not use cotton tipped 
swabs or buds as they are likely to leave fiber 
residue.

5.  Clean the two windows on the electronics 

assembly using the procedure described above.  

When the mirror assembly and electronics 

assembly are thoroughly dry, proceed with 

re-assembly.

6.  Slide a new hydrophobic filter over the end of the 

mirror assembly, then carefully seat the mirror 

assembly and hydrophobic filter against the base 

of the electronics assembly.  Be careful not to 

crumple or fold the hydrophobic filter.

7.  Tighten the two captive screws on the top of the 

mirror assembly.  Tighten the screws evenly.  Do 

not over-tighten (6 inch-pounds recommended).

8.  Slide the sample draw cup back onto the 

PIRVOLHV.

9.  Check for leaks around the PIRVOLHV sample 

draw cup.

Содержание 95-8632

Страница 1: ...Instructions 95 8632 Volumetric Infrared Hydrocarbon Gas Process Monitor for Heavy Hydrocarbons PIRVOLHV 3 1 Rev 1 12 95 8632 ...

Страница 2: ...ng Orientation 5 Termination Boxes PIRTB 5 General Wiring Requirements 5 Wiring Procedure 6 Detector Separation Optional 7 STARTUP PROCEDURE 8 TROUBLESHOOTING 8 MAINTENANCE 8 Disassembly and Cleaning Procedure 9 Calibration 10 Replacement parts 11 DEVICE REPAIR AND RETURN 11 ORDERING INFORMATION 12 APPENDIX A csa approval 13 APPENDIX B atex ce approval 14 APPENDIX C IECEx APPROVAL 15 ...

Страница 3: ...or sample extraction system PIRVOLHV is CSA certified for use in Class I Division 1 Groups B C and D and Class I Division 2 Groups A B C and D areas FEATURES Routine calibration not required Continuous self test automatically indicates a fault or fouled optics condition Performs well in the presence of high concentrations or constant background levels of hydrocarbons and in oxygen depleted atmosph...

Страница 4: ...nodized or 316 stainless steel SAMPLE DRAW CUP Material Aluminum or stainless steel Fittings Nickel plated Brass Tubing Size 1 4 inch O D DIMENSIONS See Figure 3 for dimensions of the PIRVOLHV and Figure 4 for dimensions of the PointWatch Termination Box Model PIRTB SHIPPING WEIGHT Approximate Aluminum 1 9 pounds 0 9 kilogram Stainless Steel 5 0 pounds 2 3 kilograms Table 1 Current Loop Output Lev...

Страница 5: ... 3 2 78 5 8 147 B2029 Figure 3 PIRVOL Dimensions in Inches mm Figure 4 PIRTB Dimensions in Inches cm 3 46 8 8 4 7 11 9 2 7 6 9 5 2 13 2 5 86 14 9 6 57 16 7 1 28 3 3 A2307 3 77 9 6 1 28 3 3 3 46 8 8 4 7 11 9 2 7 6 9 5 2 13 2 5 86 14 9 C2281 ...

Страница 6: ...quired it is recommended to utilize a scalable display device with a clearly marked unit of measurement in by volume Under no circumstances should a LFL unit of measure display module be used to display the PIRVOLHV signal output Operating Modes Warmup When power is applied to the PIRVOLHV it enters a Warmup mode for approximately one minute in which it performs diagnostic checks and allows the se...

Страница 7: ... cannot be guaranteed Be certain that all wiring complies with applicable regulations relating to the installation of electrical equipment in a hazardous area If in doubt consult the authority having jurisdiction before wiring the system The use of shielded cable in conduit or shielded armored cable is recommended for optimum RFI EMI protection In applications where the wiring cable is installed i...

Страница 8: ...common White lead 4 to 20 mA signal output Yellow lead Calibration input Green lead Chassis ground If the calibration wire yellow lead is not being used do not connect this wire to ground Trim excess length and insulate wire so no shorting can occur Figure 6 Tall Cover PIRTB with Window Figure 7 Short Cover PIRTB Figure 9 Typical PIRVOL Wiring with PIRTB PIRTB TERMINATION BOX D1756 PIRVOL PROCESS ...

Страница 9: ... from the power supply to the PIRVOLHV or to the transmitter and then to the PIRVOLHV to ensure that the power requirements are met Mounting and Connecting Procedure for Detector Separation The PIRTB should be mounted to a solid vibration free wall or post The termination box should be electrically connected to earth ground 1 Lubricate the sensor threads with low vapor pressure silicone grease the...

Страница 10: ...only silicone based grease not hydrocarbon based grease when lubricating threads on the PIRVOLHV and associated termination box A suitable grease is listed under Spare Parts in the Ordering Information section of this manual It is recommended to keep spare IR modules See Spare Parts section for field replacement in the event of a malfunction Table 2 Troubleshooting Table Current Level Status Corre...

Страница 11: ...IRVOLHV by pulling it straight off the electronics assembly no tools needed See Figure 12 2 Perform a visual inspection of the sample draw cup and associated plumbing and clean as required Also inspect the two O rings inside the sample draw cup and clean lubricate or replace as required 3 Loosen the two captive screws on top of the mirror assembly Figure 12 and remove the mirror assembly and hydro...

Страница 12: ...alibration Procedure To perform a zero calibration of the PIRVOLHV follow the sequence described in Table 3 NOTE The calibration sequence is initiated by momentarily connecting the calibration lead to the negative lead common of the power supply using either the Cal Magnet or an external switch If the Det Tronics PIRTB with magnetic Cal Switch is being used this is accomplished by holding the Cal ...

Страница 13: ...aging Return all equipment transportation prepaid to the factory in Minneapolis NOTE It is highly recommended that a complete spare be kept on hand for field replacement to ensure continuous protection Table 3 Calibration Sequence Description Current LED Operator Action Normal operation no gas present 4 0 mA Off If the possibility of background gases exists purge the sensor with clean air to ensur...

Страница 14: ...uminum IP66 101197 003 Stop plug M25 stainless steel IP66 102804 001 Reducer M25 to M20 aluminum 102804 003 Reducer M25 to M20 stainless steel Spare Parts Part Number Description 007128 001 Aluminum Sample Draw Cup with O Rings and Fittings 007128 003 Stainless Steel Sample Draw Cup with O Rings and Fittings 007128 002 O Ring Kit UL 006297 003 Hydrophobic Filter 006680 001 Silicone Grease Assistan...

Страница 15: ...ROL EQUIPMENT For Hazardous Locations Class I Div 1 Groups B C D T5 Class I Div 2 Groups A B C D T3C PointWatch Termination Box Model PIRTB CLASS 4828 01 SIGNAL APPLIANCES Combustible Gas Detection Instruments For Hazardous Locations Class I Div 1 Groups B C D T6 Class I Div 2 Groups A B C D T6 ...

Страница 16: ...VOL with the flying leads is to be screwed into a suitable certified flameproof enclosure with an Ingress Protection rating of at least IP66 CE MARK The PIRVOLHV Process Gas Monitor was tested and found to be compliant with EN50270 when wired in conduit or with shielded cable All screen drains shall be terminated to the chassis PointWatch Termination Box Model PIRTB 0539 FM APPROVED II 2 G Ex d II...

Страница 17: ... rating IP66 the back end of the Process Gas Monitor Model PIRVOL with the flying leads is to be screwed into a suitable certified flameproof enclosure with an Ingress Protection rating of at least IP66 PointWatch Termination Box Model PIRTB IECEx ULD 10 0002 Ex d IIC T5 T6 Gb T6 Tamb 55 C to 60 C T5 Tamb 55 C to 75 C IP66 IEC Standards IEC 60079 0 2007 IEC 60079 1 2007 IEC 60529 2 1 ed Corr 1 200...

Страница 18: ...ks of others Copyright Detector Electronics Corporation 2012 All rights reserved Detector Electronics Corporation 6901 West 110th Street Minneapolis MN 55438 USA T 952 941 5665 or 800 765 3473 F 952 829 8750 W http www det tronics com E det tronics det tronics com X3301 Multispectrum IR Flame Detector PointWatch Eclipse IR Combustible Gas Detector Eagle Quantum Premier Safety System FlexVu Univers...

Отзывы: