Despatch/Watlow 1500 Control Instruction Manual

Procedure for Programming HAST Profile (Cont.)

16. At Step 10 the display will read 10 SP, press enter.

S1 is displayed, use arrow keys to 120, press enter
S2 is displayed, use arrow keys to 85, press enter
E1 is displayed, press FCTN
E2 is displayed, press FCTN
HR is displayed, press FCTN
MN is displayed, use arrow keys to 31, press enter
SC is displayed, use arrow keys to 00, press enter
TIME is displayed 
NX is displayed, use arrow keys to 0011, press enter

17. At Step 11 the display will read 11 SP, press FCTN until

WT appears, press enter.

Press FCTN until W2 is displayed, use arrow keys to

0084, press enter

Press FCTN until NX is displayed, use arrow keys to

0012, press enter

18. At Step 12 the display will read 12 SP, press enter.

S1 is displayed, press FCTN
S2 is displayed, press FCTN
E1 is displayed, press FCTN
E2 is displayed, press FCTN
HR is displayed, press FCTN
MN is displayed, use arrow keys to 30, press enter
SC is displayed, use arrow keys to 00, press enter
TIME is displayed 
NX is displayed, use arrow keys to 0013, press enter

Содержание Watlow 1500 Control

Страница 1: ...P N 260786 Rev 8 96 E 87 U S 45 00 Instruction Manual for the Despatch Watlow 1500 Control...

Страница 2: ...equipment could cause an explosion that may result in equipment damage personal injury or possible death Dear Customer Thank you for choosing Despatch Industries We appreciate the opportunity to work...

Страница 3: ...packing installing operating and maintaining the control The APPENDIX section contains Special Instructions for operating the control instrument and a Troubleshooting Table An efficient way to learn a...

Страница 4: ...Page ii Despatch Watlow 1500 Control Instruction Manual...

Страница 5: ...rt 12 LPB Series Guarded Access Chart 13 Profile Mode 14 Change Data Mode 14 Halt Conditions 14 Blank Step 15 Set Point Step 15 Creating Files Profiles 16 Wait Steps 20 Next Step 21 Recycle Option 21...

Страница 6: ...ure 2 DIP Switch Location and Orientation 36 List of Tables Table 1 Watlow 1500 Displays 4 Table 2 Watlow 1500 Keys 5 Table 3 Watlow 1500 LEDs 5 Table 4 GA0001 and 0002 PID Tuning Parameters 10 Table...

Страница 7: ...dity from a variety of inputs The controller is easy to use Control functions alarm settings and other parameters are easily entered through the front keypad Limited data can be protected from unautho...

Страница 8: ...Page 2 Despatch Watlow 1500 Control Instruction Manual...

Страница 9: ...fans the following sequence will automatically take place 1 The controller will power up 2 The controller MNTR DATA LED will light up and the alarm indicator A1 will be flashing in the FUNCTION displa...

Страница 10: ...Lowest significant event is left justified instead of right justified as when events are shown in data display Step Shows which step the programmer is currently pointed to Function Describes the funct...

Страница 11: ...nge data and change parameter modes Increases and decreases the value in the DATA display A light touch changes the value by one Holding the key down causes the DATA display to increase rapidly Table...

Страница 12: ...ure a Press the FCTN key until C1 is signaled in the FUNCTION display b Update the dry bulb temperature by using up down arrow keys c Press enter when the desired set point is signaled in the data dis...

Страница 13: ...mage If a dry air system and an electronic humidity sensor are installed in lieu of a wet bulb sensor the optional dry air event may be outside the range of 4EC to 98EC a Press the FCTN key C2 is show...

Страница 14: ...wing GA 0000 To set CH 1 and CH 2 alarm limits GA 0001 To set CH 1 PID tune parameters GA 0002 To set CH 2 PID tune parameters GA 0006 To set set point scroll limits optional lockout and option recycl...

Страница 15: ...HR in the FUNCTION display 3 Enter the real time of day beginning with hours The CHG PARA programming prompts are listed below in the order they appear a Enter the real time of day 24 hour basis by se...

Страница 16: ...ameters Parameters FUNCTION Display Description Upper Process Type Alarm UP If you do not want this alarm set to highest value Lower Process Type Alarm LP If you do not want this alarm set to lowest v...

Страница 17: ...0 to deactivate the recycle option Temp Temp 2 channel temperature control TT This prompt will not appear on thermocou ple units Select C 0000 for Temp RH control C 0001 or Temp Temp control RS 422 A...

Страница 18: ...and 0009 1 RS Ch 1 Reset 0010 1 RT Ch 1 Rate 0000 1 Rb Ch 1 Rate Band 0001 1 CT Ch 1 Cycle Time 0003 1 db Ch 1 Dead Band 0001 1 CA Ch 1 Calibration Adjust 0000 Dual PID 1C Pb Ch 1 Prop Band 0009 1C RS...

Страница 19: ...Lower Process Alarm 001 C2 Ud Ch 2 Upper Deviation Alarm 101 C2 Ld Ch 2 Lower Deviation Alarm 101 0005 RC Recycle 0000 C1 Pb Ch 1 Prop Band 0003 C1 RS Ch 1 Reset 0005 C1 RT Ch 1 Rate 0001 C1 Rb Ch 1...

Страница 20: ...ssor There are three step types C Set Point SP C Jump Loop JL C Wait WT Halt Conditions The Watlow 1500 can HALT in three ways when the C processor encounters a Blank Step in the program C RUN HALT ke...

Страница 21: ...t stop until reaching steps 47 51 required for cooldown and shutdown Set Point Step The existing start conditions for each step depend upon the programmed step previously performed by the Watlow 1500...

Страница 22: ...s the FCTN key until SP appears in the function display 3 Press the ENTER key S1 will appear in the FUNCTION display and 0000 will appear in the DATA display 4 Use the FCTN key to select the variable...

Страница 23: ...key and the key to enter the JS Jump Step parameter OR Use the FCTN key to advance to the next prompt without entering a value b Press the ENTER key c Use the key and the key to enter the JC Jump Coun...

Страница 24: ...wait condition for the CH 1 set point can be entered b Use the key and the key to enter the wait condition OR Use the FCTN key to advance to the next prompt without entering a value c Press the ENTER...

Страница 25: ...e 1 or 0 entry for Events 4 through 1 from left to right Press ENTER NOTE If you ve set DIP Switch 1 ON then there are only three events available for each channel In that case each Event 4 is an alar...

Страница 26: ...H WM or WS is shown you could end up waiting for midnight when you wanted wait for time ignored 2 Table 8 Wait Steps Wait Steps The Wait WT submode allows the Watlow 1500 to wait for C Channel 1 and o...

Страница 27: ...d information If there is data in the subsequent step the processor will move to the first prompt in that submode If you want a different step from NX select the number of the step you want in the DAT...

Страница 28: ...ber of steps in the program may be as many as 51 We suggest that you write it out on copies of the charts in this section This will enable you to program the Series 1500 quickly and without mistakes T...

Страница 29: ...Programming Work Sheet Name Date Step No Set Point Events Duration Comments CH 1 EC or EF S1 CH 2 or E S2 1 ON E1 4321 0 OFF E2 4321 HR MN SC 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 2...

Страница 30: ...mpression 16 Blank step 17 18 19 20 130 10 Ramp C1 to 130EC 1EC min first not RH 21 85 05 Ram RH after temperature has been reached 22 W2 84 Wait for near set conditions 23 0000 00 30 Hold for 30 min...

Страница 31: ...isplayed use arrow keys to 70 press enter E1 is displayed use arrow keys to 0000 press enter E2 is displayed use arrow keys to 0001 press enter HR is displayed use arrow keys to 00 press enter MN is d...

Страница 32: ...until SC is displayed use arrow keys to 01 press enter TIME is displayed NX is displayed use arrow keys to 0005 press enter 11 At Step 5 the display will read 05 SP press enter S1 is displayed use arr...

Страница 33: ...er 14 At Step 8 the display will read 08 SP press enter S1 is displayed use arrow keys to 120 press enter S2 is displayed use arrow keys to 59 press enter E1 is displayed press FCTN E2 is displayed us...

Страница 34: ...is displayed NX is displayed use arrow keys to 0011 press enter 17 At Step 11 the display will read 11 SP press FCTN until WT appears press enter Press FCTN until W2 is displayed use arrow keys to 008...

Страница 35: ...ss enter JC is displayed use arrow keys to 0001 press enter Press FCTN until NX is displayed use arrow keys to 0015 press enter 21 At Step 15 the display will read 15 SP Press FCTN until JL appears pr...

Страница 36: ...ress FCTN SC is displayed use arrow keys to 01 press enter TIME is displayed NX is displayed use arrow keys to 0050 press enter 25 At Step 50 the display will read 50 SP press FCTN until WT appears pr...

Страница 37: ...noid 0 Open 1 Closed Drain Solenoid 0 Closed 1 Open HAST Chamber Programming Instructions and Definitions C1 Channel 1 Temperature EC C2 Channel 2 Relative Humidity RH SP Set Points S1 Set Temperature...

Страница 38: ...ld for 30 minutes 6 S1 is changed to 10EC This is ramped over a 1 second period so it is essentially a step change Note Event 1 1 is turned off To protect the equipment humidity control generally need...

Страница 39: ...0 30 00 Ramp to 25EC in 30 minutes 5 00 30 00 Hold 25EC for 30 minutes 6 10 0001 00 00 01 Step Ramp to 10 Humid event OFF 7 W1 8 Wait for 8EC ALL OTHER W S BLANK 8 00 30 00 Hold 10 for 30 minutes 9 JS...

Страница 40: ...Program To clear an entire program perform the following steps 1 Press MODE key until CHG PARA LED lights 2 Press FCTN key until GA shows in FUNCTION display 3 Press the key until DATA display shows 0...

Страница 41: ...wer Deviation Alarm 0001 1H Pb Ch 1 Prop Band 1H RS Ch 1 Reset 1H RT Ch 1 Rate 1H Rb Ch 1 Rate Band 1H CT Ch 1 Cycle Time 1 db Ch 1 Dead Band 1 CA Ch 1 Calibration Adjust Dual PID 1C Pb Ch 1 Prop Band...

Страница 42: ...emory 2 A Cold start is a Clean startup condition All MEMORY IS DELETED and the controller will return to default settings 3 Upon completion of Cold start return the control to Warm start otherwise th...

Страница 43: ...ry use only Must be in the ON position 6 Cold start on power up Memory cleared parameters set to default values Warm start on power up Programmed values are retained for all parameters 7 EC function a...

Отзывы: