Desoutter DPX-170 Скачать руководство пользователя страница 6

4

Automatic
Cleaning

Equipment:

Automatic Washer/Disinfector capable of meeting relevant national and
international cleaning and disinfection standards i.e. BS2745 and
HTM2030; Neutral pH Enzymatic detergent such as HAMO Liquid 52 

.

Method:

1. 

Large contaminant deposits should be removed manually using the
method described in Manual Cleaning:- item 1. Pay particular attention
to recesses, blind holes, chucks, clamps and cannulations.

2. 

Place the handpiece, attachments and accessories onto the wire
basket. Ensure all items are separated and washing caps are securely
fitted to the handpiece and air supply hose.

Note:

The placement of item in automatic washer/disinfector baskets can be a
critical factor in achieving effective cleaning. Selection of the basket type
and position of the items to be cleaned should be done by suitably trained
personnel in accordance with the manufacturer’s instructions for the
washer/disinfector.

3. 

Follow manufacturers loading instructions and select the appropriate
cycle recommended.

The cycle should include:

§

 

Pressurized cold water rinse (maximum 35 

°

C).

§

 

Hot water wash (minimum 55 

°

C) using a neutral pH enzymatic

detergent.

§

 

Warm water rinse.

§

 

Disinfection rinse (minimum 80 

°

C for 1 minutes).

§

 

Drying cycle

4. 

Remove disinfected instruments from the washer/disinfector to a clean
area. Remove the washing caps.

5. 

Visually inspect each item and verify the cleaning process is complete
and all contaminants have been removed in accordance with local
reprocessing guidelines.

Disinfection:

Thermal disinfection is recommended and included in the automatic
washer/disinfection cycle. See above.

Maintenance:

Lubrication of the motor is not essential but will enhance the life and
performance of the instrument.

1.  Apply 3 drops of surgical instrument oil (part no. 30982- Non-sterile)

into the instrument air inlet after cleaning.

2.  Connect the instrument to an air supply and run for at least 20 seconds

to ensure adequate dispersal of lubricant.

3.  Lubricate collets, chucks and hose connectors using mineral oil such as

Blitz

.

Inspection and
Function
Testing:

Ensure the handpiece, attachments and non-cutting accessories are in
good working order. Note any unusual sounds, vibrations or operating
speeds. If operating difficulties are experienced and are not already

Содержание DPX-170

Страница 1: ...Operating Instructions Pneumatic Oscillating Saw International English Sternum Saw DPX 170...

Страница 2: ......

Страница 3: ...Safety Instructions 2 Product Description 2 Air Supply 2 Reprocessing Instructions 3 Air Supply Hose 6 Controls 7 Blade Clamp Operation 8 Conditions for Transport and Storage 8 Explanation of Symbols...

Страница 4: ...AFE position when changing accessories or when not in use Always use eye protection when cutting to prevent injury from flying debris Do not allow loose articles to be caught by the moving parts of th...

Страница 5: ...rried out where automatic washer disinfection is not available It should be conducted in a dedicated area by trained personnel wearing protective clothing e g gloves waterproof apron and goggles or vi...

Страница 6: ...lude Pressurized cold water rinse maximum 35 C Hot water wash minimum 55 C using a neutral pH enzymatic detergent Warm water rinse Disinfection rinse minimum 80 C for 1 minutes Drying cycle 4 Remove d...

Страница 7: ...s in accordance with EN868 is recommended to preserve sterility The material should present a barrier to micro organisms and particulate contamination Additional Information Automated cleaning was val...

Страница 8: ...The connector will come to a stop the hose is now connected Pull gently on the hose to check that the hose is securely connected to the instrument To remove the hose Hold the instrument and grip the h...

Страница 9: ...13 22 S81 104 0 5 1 0 1 5 42 51 S81 105 0 5 1 0 1 8 23 32 S81 106 0 5 1 0 1 4 15 25 For full details of suitable saw blades refer to Stericut data sheets Controls Lever The lever control has a progre...

Страница 10: ...lined hole in the blade over the spline on the shaft Replace the clamp nut and tighten Removal Using the spanners provided remove the clamp nut on the end of the oscillating shaft in an anti clockwise...

Страница 11: ...ufficient lubrication Refer to Lubrication section of manual Insufficient air supply pressure and or flow rate Check air supply pressure at regulator bottled air only Check hose for possible restricti...

Страница 12: ...nd send to the address below Please ensure that the instrument has been properly decontaminated and sterilised Guarantee Liability De Soutter Medical guarantees all instruments attachments and accesso...

Отзывы: