Operating Instructions
PneumaticOscillating Saw
International
English
Sternum Saw
DPX–170
Страница 1: ...Operating Instructions Pneumatic Oscillating Saw International English Sternum Saw DPX 170...
Страница 2: ......
Страница 3: ...Safety Instructions 2 Product Description 2 Air Supply 2 Reprocessing Instructions 3 Air Supply Hose 6 Controls 7 Blade Clamp Operation 8 Conditions for Transport and Storage 8 Explanation of Symbols...
Страница 4: ...AFE position when changing accessories or when not in use Always use eye protection when cutting to prevent injury from flying debris Do not allow loose articles to be caught by the moving parts of th...
Страница 5: ...rried out where automatic washer disinfection is not available It should be conducted in a dedicated area by trained personnel wearing protective clothing e g gloves waterproof apron and goggles or vi...
Страница 6: ...lude Pressurized cold water rinse maximum 35 C Hot water wash minimum 55 C using a neutral pH enzymatic detergent Warm water rinse Disinfection rinse minimum 80 C for 1 minutes Drying cycle 4 Remove d...
Страница 7: ...s in accordance with EN868 is recommended to preserve sterility The material should present a barrier to micro organisms and particulate contamination Additional Information Automated cleaning was val...
Страница 8: ...The connector will come to a stop the hose is now connected Pull gently on the hose to check that the hose is securely connected to the instrument To remove the hose Hold the instrument and grip the h...
Страница 9: ...13 22 S81 104 0 5 1 0 1 5 42 51 S81 105 0 5 1 0 1 8 23 32 S81 106 0 5 1 0 1 4 15 25 For full details of suitable saw blades refer to Stericut data sheets Controls Lever The lever control has a progre...
Страница 10: ...lined hole in the blade over the spline on the shaft Replace the clamp nut and tighten Removal Using the spanners provided remove the clamp nut on the end of the oscillating shaft in an anti clockwise...
Страница 11: ...ufficient lubrication Refer to Lubrication section of manual Insufficient air supply pressure and or flow rate Check air supply pressure at regulator bottled air only Check hose for possible restricti...
Страница 12: ...nd send to the address below Please ensure that the instrument has been properly decontaminated and sterilised Guarantee Liability De Soutter Medical guarantees all instruments attachments and accesso...