TB-6542
Page 4 of 5
Revision 10/04
LOW LIMIT
CALIBRATION
RESISTANCE
LOW LIMIT FAIL LOW
RESISTANCE
[1M - (1M x 10%)]
LOW LIMIT PASS
SAFE RESISTANCE
[1M + (1M x 10%)]
1M
Ω
900K
Ω
1.1M
Ω
HIGH LIMIT
CALIBRATION
RESISTANCE
HIGH LIMIT PASS SAFE
RESISTANCE
[15M
Ω
- (15M
Ω
x 10%)]
HIGH LIMIT FAIL HIGH
RESISTANCE
[15M
Ω
+ (15M
Ω
x 10%)]
15M
Ω
13.5M
Ω
16.5M
Ω
C.)
The table below specifies which numbered resistances on the circuit board
correspond to the optional low limit and the optional high limit. Solder the appropriate
resistors in these places.
Optional low limit
Designated resistors on
circuit board
Low limit fail low
R1, R1A (*connected in
series)
Low limit pass safe
R2, R2A (*connected in
series)
Optional high limit
Designated resistors on circuit
board
High limit pass safe
R9, R9A (*connected in series)
High limit fail high
R10, R10A, R10B (*connected in
series)
NOTE
: Two (three for position 10) resistor locations are connected in series are
provided in case resistors need to be added to achieve the desired total resistance.
Solder a shorting wire across the resistor locations that are not used.
D.) Assemble the unit and mark the appropriate places on the label for the optional
limits.
E.) It is recommended to turn the rotary switch to the optional positions and measure
the resistance from the tip to the body of the plug with an ohmmeter to verify the
resistances.
ohmmeter
MAINTENANCE
:
Wipe the plug periodically with alcohol.