Desco 60506 Скачать руководство пользователя страница 5

TB-3043

 Page 5 of 6

© 2019 DESCO INDUSTRIES INC

Employee Owned

DESCO WEST 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178

DESCO EAST 

- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: 

Desco.com

TP1

+16.0
-25.0

12"

12"

12"

12"

12"

TP4

+9.5
-12.0

TP7

+11.5
-15.0

TP10

+15.5
-21.0

TP2

+1.3
-1.8

TP5

+3.0
-4.1

TP8

+5.0
-6.5

TP11

+7.0
-10.0

TP3

+38.0
-59.0

TP6

+11.0
-16.0

TP9

+10.0
-12.5

TP12

+11.0
-14.0

High Output 

Bench Top Ionizer

Air Flow

Charged Plate

12"

Figure 6.  Neutralization (Discharge) Times at 100VAC, 50Hz input

Calibration

On a regular interval, most users will clean emitter pins and 

calibrate the ionizer.  Per ESD TR 53 section 5.3.6.7.1 “The 

best practice is to measure the offset voltage and discharge 

times, clean the unit, including emitter points and air filters 

if present, offset voltage to zero (if adjustable), and then 

repeat offset voltage and discharge time testing.  If the unit 

does not meet offset voltage specifications or minimum 

established discharge time limits, further service is indicated. 

Manufacturers should provide details on service procedures 

and typical service intervals.”
Most companies will assign a number or otherwise 

identify each ionizer and setup a compliance Verification / 

Maintenance / Calibration schedule.  If the ionizers all test 

good, the data can justify lengthening the calibration period.  

If ionizers require adjustment, the calibration period should 

be shortened.  Although ESD TR53 does not advise a test 

frequency, JESDD625-A (Revision of EIA-625) recommends 

ionizers be tested semiannually, noting to use “S3.1 except 

the number of measurement points and locations may be 

selected based on the application.”

NOTE: A charged plate analyzer or monitor should be 

used in order to properly calibrate the Desco High Output 

Benchtop Ionizer.  EMIT offers the 

50571

 Charged Plate 

Analyzer.

1.  Properly setup the ionizer as described in the installation 

 procedure.
2.  Turn the unit ON and set the fan speed to High.
3.  Position the charged plate analyzer 12 inches directly in 

  front of the ionizer.
4.  The balance (offset voltage) should be within 0 and ±2 

  volts.  To increase the output in a positive direction, turn 

  the Balance potentiometer in a clockwise direction.  To 

  increase the output in a negative direction, turn the 

  Balance potentiometer in a counter-clockwise direction.
5.  Test the neutralization (discharge) time by applying a 

  ±1,000 volt charge on the charged plate analyzer.  The 

  neutralization (discharge) time should be less than 2 

  seconds.  See figures 5 and 6 for typical discharge times.  

  The required limit per ANSI/ESD S20.20 is “user defined”.
6.  Test the ionizer’s alarm by shorting its two fan grills 

  located on the front (see Figure 7).  The alarm should 

  sound, and the STATUS LED should illuminate red.

Содержание 60506

Страница 1: ...rrounding air When an electrostatic charge is present on objects in the work environment it will be neutralized by attracting opposite polarity charges from the ionized air Note that ionization systems should not be used as a primary means of charge control on conductors or people Reference EN 61340 5 2 1 clause 5 2 9 September 2019 The primary method of static charge control is direct connection ...

Страница 2: ...nit OFF and unplug the power cord A B C E D Figure 2 High Output Benchtop Ionizer features and components A Power Cord Connection Connect the power cord here B Power Switch Toggle the switch to the left to turn the ionizer OFF Toggle the switch to the right to turn the ionizer ON C Fan Speed Switch Toggle the switch to the left to set the fan speed to Low Toggle the switch to the middle to set the...

Страница 3: ...will continuously sound The user must then reset the unit by turning it OFF and back ON WARNING RISK OF ELECTRIC SHOCK THESE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY DO NOT PERFORM ANY SERVICING OF INTERNAL PARTS UNLESS YOU ARE QUALIFIED TO DO SO NOTE The AC power cord MUST always be disconnected before the unit is disassembled The input voltage may be verified or reset by re...

Страница 4: ...ments recorded in a factory ambient environment Per ANSI ESD S20 20 the test method for Product Qualification test is ANSI ESD STM3 1 and for Compliance Verification is ESD TR53 which advises Measurements should be made at the location where ESD sensitive items are to be ionized A larger area may require additional ionizers Per S20 20 the required limit for ionizer discharge time is user defined U...

Страница 5: ... the data can justify lengthening the calibration period If ionizers require adjustment the calibration period should be shortened Although ESD TR53 does not advise a test frequency JESDD625 A Revision of EIA 625 recommends ionizers be tested semiannually noting to use S3 1 except the number of measurement points and locations may be selected based on the application NOTE A charged plate analyzer ...

Страница 6: ... W Dimensions including stand 9 5 H x 6 0 W x 3 1 D 241 mm x 152 mm x 79 mm Weight 4 5 lbs 2 0 kg Balance Offset Voltage at 12 3 V typical 25 V maximum Neutralization Discharge Time at 12 2 seconds Airflow Three speed fan 50 100 CFM Emitter Points 050 diameter tungsten Ion Emission Steady state DC with sense feedback Fuse 250 mA slow blow High Voltage Power Supply 5 5 kVDC nominal Ozone 0 05 ppm E...

Страница 7: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Desco 60505 60507 ...

Отзывы: