Neutralization (Discharge) Times

All data was taken with the fan speed set to high.  All time measurements are in seconds.
NOTE: Discharge times in seconds are representative only and are not a guarantee.  Discharge times are actual 

measurements recorded in a factory ambient environment.
Per ANSI/ESD S20.20, the test method for Product Qualification test is ANSI/ESD STM3.1, and for Compliance 

Verification is ESD TR53 which advises “Measurements should be made at the location where ESD sensitive items 

are to be ionized.”  A larger area may require additional ionizers.  Per S20.20 the required limit for ionizer discharge 

time is user defined.  Use Table to determine the number of ionizers to achieve ionization of area to be neutralized 

to meet your company’s ESD control plan specified discharge times.
The comparative efficiency of bench top ionizers is determined by a standard test published by the ESD 

Association: ANSI/ESD STM 3.1.  Typical positive and negative decay times (1000V - 100V) measured using this 

standard are shown in the figures below.  The performance of the ionizer was measured with the unit positioned as 

shown, with the fan speed on high and without a filter.

TP1

>60(+)
>60(-)

12"

12"

12"

12"

12"

TP4

>60(+)
>60(-)

TP7

28.0(+)
32.0(-)

TP10

18.0(+)
20.0(-)

TP2

1.2(+)
1.4(-)

TP5

1.5(+)
2.0(-)

TP8

2.3(+)
3.0(-)

TP11

3.1(+)
4.0(-)

TP3

>60(+)
>60(-)

TP6

>60(+)
>60(-)

TP9

20.0(+)
23.0(-)

TP12

15.0(+)
20.0(-)

Bench Top

Blower Ionizer

Air Flow

Charged Plate

12"

Figure 7.  Neutralization (Discharge) Times of item 

19500

 at 120VAC, 60Hz input

Содержание 19520

Страница 1: ...Ionization is not effective on items that have large capacitance like people and carts however ionizers should be considered as a method for charge neutralization in cases where grounding cannot be achieved Air ionization can neutralize the static charge on insulated and isolated objects by producing separate charges in the molecules of the gases of the surrounding air When an electrostatic charge...

Страница 2: ...ments with more contaminants METHOD 1 Annually To remove heavier accumulations of dirt and dust use clean compressed air Disconnect power to the ionizer With one hand firmly on top of the ionizer aim com pressed air through the front grill directly at the pins A few passes back and forth with the nozzle is sufficient Figure 3 Wall or shelf mounting of the ionizer Set both front panel rocker switch...

Страница 3: ... and negative high voltage usually at the power line frequency 50 60 Hz The ionizer s unique Faraday Balance System monitors ion output and automatically adjusts the unit to reach an equilibrium at low offset voltage or balanced ion flow The ionizer is designed to achieve balance under extreme conditions and is uniquely able to maintain balance in most any environment Ions are emitted in an effici...

Страница 4: ... may require additional ionizers Per S20 20 the required limit for ionizer discharge time is user defined Use Table to determine the number of ionizers to achieve ionization of area to be neutralized to meet your company s ESD control plan specified discharge times The comparative efficiency of bench top ionizers is determined by a standard test published by the ESD Association ANSI ESD STM 3 1 Ty...

Страница 5: ...TP8 2 8 3 6 TP11 3 7 4 9 TP3 60 60 TP6 60 60 TP9 35 0 41 0 TP12 21 0 26 0 Bench Top Blower Ionizer Air Flow Charged Plate 12 Figure 9 Neutralization Discharge Times of item 19520 at 230VAC 50Hz input TP1 60 60 12 12 12 12 12 TP4 60 60 TP7 38 0 46 0 TP10 21 0 28 0 TP2 1 4 1 6 TP5 2 0 2 5 TP8 2 9 3 7 TP11 3 9 4 9 TP3 60 60 TP6 60 60 TP9 35 0 40 0 TP12 19 0 25 0 Bench Top Blower Ionizer Air Flow Char...

Страница 6: ...19500 120 VAC 50 60 Hz 3 0 Amps 19520 220 VAC 50 60 Hz 1 5 Amps Fuse Type 19500 Slow blow 5 amp 19520 Slow blow 3 amp Output 4 5kV at 50 60 nano amps Air Flow Adjustable 117 CFM maximum Balance Voltage 5V Dimensions 11 x 9 x 7 Weight 19500 13 Pounds 19520 16 Pounds Cord 19500 Cord with three wire IEC plug ground required 19520 Not included Desco ionizers meet the ANSI ESD S20 20 minimum recommende...

Страница 7: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Desco 19500 ...

Отзывы: